Specification And Dimension Of Each Part - LG 17HQ701G-BP.AEU Owner's Manual

Flat panel digital x-ray detector
Table of Contents

Advertisement

SPECIFICATION AND DIMENSION OF EACH PART

Specifications are subject to change without notice.
The symbol ~ means alternating current, and the symbol
Specifications
Detector
Item
Model
17HQ701G
Sensor Type
a-Si TFT
Scintillator Type
CsI:TI
Total Pixel Matrix
3072 x 3072 pixels
Total Pixel Area
430.08 mm x 430.08 mm
Pixel Pitch
140 um
Effective Pixel Matrix
3060 x 3060 pixels
A/D Conversion
16 bits
Data transmission
802.11 a/b/g/n/ac Wireless LAN Standard, 150 Mbps
Wired Gigabit Ethernet Standard, 500 Mbps
Cycle time
Typ. 4.5 Sec (Wired)
Typ. 5 Sec (Wireless)
Image Transmission
Typ. 1.5 Sec (Wired)
Typ. 2.5 Sec (Wireless)
Image Save
Stores up to 200 images
Semi-dynamic mode
5fps @ 140 um (full FOV)
Energy Range
40 kVP ~ 150 kVp
MTF
Typ. 84 % @ 0.5 lp/mm
DQE
Typ. 66 % @ 0.1 lp/mm
means direct current.
Specifications
Item
Size (Width x Height x Depth) (mm)
460.0 x 460.0 x 15.6
Weight (kg)
Typ. 3.6
Window Materials
Carbon Fibre
Trigger Mode
Manual Mode
Auto Mode (Auto Exposure Detection)
Power Consumption
Typ. 38 W
Wireless
Standard:
802.11 a/b/g/n/ac compliance
Peak Mode: 867 Mbps
Frequency: 2.4 GHz / 5 GHz
Bandwidth: 20 MHz / 40 MHz / 80 MHz
MIMO: 2 x 2
Rating
24 V
Applied part
Type: BF
Location: The front side of the Detector (Effective area only)
NOTE
Maximum wireless signal rate derived from IEEE standard specifications. Actual data throughput will vary. Network
conditions and environmental factors, including volume of network traffic, building materials and construction,
and network overhead, lower actual data throughput rate.
Recommended Maximum operable distance: 2 m (From the Access Point)
Wireless antennas: The module adopts the latest 802.11ac technology. The transmitter of the module is powered by
host equipment (Detector). The antennas are 2 printed-dipole antennas.
Wireless module: 802.11 a/b/g/n/ac USB 2.0 module is implemented. It supports 2T2R (2 transmit 2 receive) MIMO
technology, which delivers throughput up to 300 Mbps.
Images can be saved by the X-ray generator while the power of the detector is turned on without connecting to a
PC. To produce images, X-ray is irradiated at intervals of more than 10 seconds. Check and load the saved images
from LG Acquisition Workstation Software.