Specification And Dimension Of Each Part - LG 14HK701G-W.AUB Owner's Manual

Flat panel digital x-ray detector
Table of Contents

Advertisement

SPECIFICATION AND DIMENSION OF EACH
PART
The product specifications are subject to change without prior notice for product improvements.
~ refers to alternating current (AC),
refers to direct current (DC).
Specifications
Detector
Item
Model
14HK701G
Sensor type
Amorphous Silicon TFT
Scintillator Type
CsI:TI
Total Pixel Matrix
2500 x 3052 pixels
Total Pixel Area
350 x 427.28 mm
Pixel Pitch
140 um
Effective Pixel Matrix
2488 x 3040 pixels
A/D Conversion
16 bits
Data transmission
802.11 a/b/g/n/ac Wireless LAN Standard, 150 Mbps
Wired Gigabit Ethernet Standard, 500 Mbps
Cycle time
Typ. 8 Sec (Wired)
Typ. 11 Sec (Wireless)
Image Transmission
Typ. 2 Sec (Wired)
Typ. 2.5 Sec (Wireless)
Image Storage
Stores up to 200 images
Energy range
40 kVp ~ 150 kVp
MTF
Typ. 89 % at 0.5 lp/mm
DQE
Typ. 72 % at 0.1 lp/mm
Specification
Item
Size
384.0 x 460.0 x 15.2 mm
(Width x Height x Depth)
(15.1 x 18.1 x 0.5 inch)
Weight
Typ. 2.95 kg (6.5 lbs)
Window material
Carbon Fiber
Trigger mode
Manual Mode
Auto Mode (Auto Exposure Detection)
Power consumption
Typ. 28 W
Wireless
Standard:
802.11 a/b/g/n/ac compliance
Peak mode: 867 Mbps
Frequency: 2.4 GHz / 5 GHz
Bandwidth: 20 MHz / 40 MHz / 80 MHz
MIMO: 2X2
Rating
24 V
2.1 A
Applied part
Type: BF
Location: The front side of the Detector (Effective area only).
NOTE
Maximum wireless signal rate derived from IEEE standard specifications. Actual data throughput will vary.
Network conditions and environmental factors, including volume of network traffic, building materials and
construction, and network overhead, lower actual data throughput rate.
Recommended Maximum operable distance: 2 m (6.5 ft) (From the Access Point)
Wireless antennas: The module adopts the latest 802.11ac technology. The transmitter of the module is
powered by host equipment (Detector). The antennas are 2 printed-dipole antennas.
Wireless module: 802.11 a/b/g/n/ac USB2.0 module is implemented. It supports 2T2R (2 transmit 2 receive)
MIMO technology, which delivers throughput up to 300 Mbps.
Images can be saved by the X-ray generator while the power of the detector is turned on without connecting
to a PC. To produce images, X-ray is irradiated at intervals of more than 10 seconds. Check and load the saved
images from
Specification
.
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

14hk701g-w14hk701g-wp.awp

Table of Contents