LG 14HK701G-W.AUB Owner's Manual page 22

Flat panel digital x-ray detector
Table of Contents

Advertisement

Trigger Cable
Trigger Cable is connected between control box and X-ray Generator, and used only for manual mode, not auto mode.
DXD 1
DXD 2
Ethernet
Sync
AC-IN
T4L 250V
No.
Labeling of pins color
1
Red
2
Brown
3
Blue
4
Orange
5
Yellow
6
Green
7
Black
8
Gray
9
Violet
NOTE
Trigger Cable and X-ray Generator connection is implemented by expert of X-ray system manufacture. Description of each pin is common language of this industry.
22
Description
Power : X-ray Generator Supply Voltage (DC 12V ~24V)
Expose signal from Generator to Control Box
Cancel REQ Ground
Prepare Acknowledge signal from Control Box to Generator
Expose Acknowledge signal from Control Box to Generator
Prepare signal from Generator to Control Box
Cancel request VCC
Prepare Acknowledge Ground
Ground of Signals
NC : No Connection
Use
Use
NC
Use
Use
Use
NC
NC
Use
X-ray
Generator