Siemens UROSKOP ACCESS Installation And Startup
Siemens UROSKOP ACCESS Installation And Startup

Siemens UROSKOP ACCESS Installation And Startup

Urodynamic interface/scan converter
Hide thumbs Also See for UROSKOP ACCESS:

Advertisement

Quick Links

K ö l l n e r
UROSKOP ACCESS
Installation and Start-up
System
Urodynamic Interface/Scan Converter
Print No.:
SPL5-330.814.01.05.02
Replaces: SPL5-330.814.01.04.02
05756130
05756122
SP
2003
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.05
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens UROSKOP ACCESS

  • Page 1 K ö l l n e r UROSKOP ACCESS Installation and Start-up System Urodynamic Interface/Scan Converter 05756122 05756130 2003 © Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Protective Conductor Test ..........17 3 _______ Changes to previous version _____________________________________ 18 Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 3 of 18 Medical Solutions 08.05...
  • Page 4: Requirements

    If an uninterruptible power supply (USV) is installed in the system, the voltage output of the USV must also be deenergized or the voltage output plug must be discon- nected. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 4 of 18 08.05...
  • Page 5 Avoid all contact with blood or other bodily fluids! Strictly observe the safety information in ARTD-002.731.37 regarding prophylactic measures against infectious diseases during service. Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 5 of 18 Medical Solutions 08.05 CS PS 24...
  • Page 6: Additionally Required Documents

    Medical Products; Safety Information; General TD00-000.860.01 Safety Notes Installation Instructions; System SPL5-330.812.01 Abbreviations Electrostatically sensitive device Imaging system FLUOROSPOT Compact Optical isolation amplifier Uninterruptible Power Supply UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 6 of 18 08.05 CS PS 24 Medical Solutions...
  • Page 7: Urodynamic Interface/Scan Converter

    Upgrade kit, optical isolation amplifier (OTV; part no. 57 56 494) Number Part numer Description 46 72 028 OTV80M2 75 58 518 OTV power cable 37 95 239 BNC video cable Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 7 of 18 Medical Solutions 08.05 CS PS 24...
  • Page 8: Installation

    (3/Fig. 1 / p. 8) und den OTV (4/Fig. 1 / p. 8) (optical isolation amplifier) in the imaging system container of the UROSKOP Access. - No OTV is installed if an endoscopic interface is present. Fig. 1: Imaging system container Pos.
  • Page 9: Scan Converter (Connections)

    Urodynamic interface Scan converter (connections) Fig. 2: Image signal splitter (connections) Fig. 3: Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 9 of 18 Medical Solutions 08.05 CS PS 24...
  • Page 10 Urodynamic interface Fig. 5: Pos. 1 BNC 1 urodynamic option Fig. 4: Urodynamic interface UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 10 of 18 08.05 CS PS 24 Medical Solutions...
  • Page 11: Installation And Wiring Of The Scan Converter

    9. Use the cable ties to secure the cables. 10. Connect the BNC cable located at the lifting base in the back wall of the device (moni- tor switchover URO). 11. Install the urodynamic interface in the back wall of the UROSKOP Access lifting base. Siemens AG SPL5-330.814.01.05.02...
  • Page 12: Programming The Scan Converter

    ASPIA imaging system. • Press the Menu key until Adjust: Screen size appears. • Press the + key. - The sub-menu is displayed. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 12 of 18 08.05 CS PS 24 Medical Solutions...
  • Page 13: Cutting Off The Menu Bar (Fluorospot Compact)

    • To save all settings, press the Menu key and the + key at the same time.. • Switch the UROSKOP Access off and back on again. Cutting off the menu bar (FLUOROSPOT Compact) NOTE This subsection refers exclusively to UROSKOP-Access systems with the FLC imaging system.
  • Page 14 3. For the scan converter model TVone CS-500A, this parameter is no longer selectable. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 14 of 18 08.05...
  • Page 15: Scan Converter Parameter Table (Fluorospot Compact Imaging System)

    2. The parameters in the menu “Manual set“ (VGA left, VGA width, VGA top/4, VGA bott/4) can vary from customer to customer, depending on the factory-set monitor settings implemented by the manufacturer in the urodynamic system. Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 15 of 18 Medical Solutions 08.05 CS PS 24...
  • Page 16: Faulty Screen Display

    Urodynamic interface Faulty screen display • To revert to the factory settings, switch off the scan converter, press and hold the + and - keys simultaneously, and restart the scan converter. UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 16 of 18 08.05...
  • Page 17: Final Work Steps

    A sudden or unexpected increase in the measured values may in- dicate a defect in the ground wire connections (ground wire or contacts) - even if the limit value of 0.2 ohms is not exceeded. Siemens AG SPL5-330.814.01.05.02 UROSKOP ACCESS Page 17 of 18 Medical Solutions 08.05 CS PS 24...
  • Page 18 “Cutting off the menu bar the scan converter“ (FLUOROSPOT Compact)“ new; “scan converter parameter table (FLUOROSPOT Compact imaging system)“ new Subsection “Final work “Protective conductor test“ new steps“ UROSKOP ACCESS SPL5-330.814.01.05.02 Siemens AG Page 18 of 18 08.05 CS PS 24 Medical Solutions...

Table of Contents