Philips PM 5775 Manual page 47

Table of Contents

Advertisement

Auxiliary pulse generator PM 5770
REPETITION TIME
DELAY
WIDTH
PULSE MODE
POLARITY
RAMP RANGE
RISE TIME
FALL TIME
DC-OFFSET
AMPLITUDE
Sampling oscilloscope PM 3400
TIME SCALE MAGN.
TIME/cm
TRIGG./SENSITIVITY
Pulse generator to be tested PM 5775 (PM 5776)
REPETITION TIME, Si
DELAY, S82, RV2
WIDTH, $3, RV3
AMPLITUDE IN 50 , S4, RV5
DC OFFSET IN 50 2, RV4
CONT. GATED, S6
SINGLE DOUBLE, $7
NORM. INV., S8
Auxiliary
pulse generator
Generator
under test
Fig. X-1. Set-up for checking the gate circuit
PM5770 / PMS775
PM5775 (PM5776)
T-piece
BNC
UG — 274/U
55
EXT.
20 ns
200 ns
CONT, SINGLE
NORM, positive (+)
<4 —50ns
fully counter-clockwise
fully counter-clockwise
OV
12V
10 ns
SYNC.
10 ns
5 ns
5 ns
+1 V
OV
GATED
SINGLE
NORM.
Sampling
oscilloscope
PM3400
Trigg out
oO
YA
YB
O
O
MA6494