HP 10430A Operating Note page 20

Miniature passive probes
Table of Contents

Advertisement

NOTE
Adjust for low-frequency
compensation before performing
high-frequency compensation or
adjustment may not be accurate.
To accomplish high-frequency compensation
proceed with the following steps:
1.
Using 50-ohm coaxial cable, connect pulse
generator directly to channel input
connector on oscilloscope. Terminate
oscilloscope into 50 ohms.
2.
Verify that both channels of oscilloscope
meet their published rise time/perturbation
specifications.