HP 10400A OPERATING NOTE
9.
Adjust pulse generator output for
oscilloscope display of approximately 6
divisions.
10. Slide off probe chassis assembly cover to
expose adjustments. Refer to figure 7 for
adjustment locations.
11. Set high-frequency compensation resistor
R1 to minimum resistance (completely
counterclockwise).
12. Set high-frequency compensation resistor
R2 to minimum resistance (completely
clockwise).
Figure 7. High-frequency Compensation
Adjustment Locations
18