HP 5308A Operating And Service Manual page 18

75 mhz timer/counter
Table of Contents

Advertisement

Model 5308A
Operation
Ratio Seli Check
TIME BASE
DISPLAY
ANNUNCIATOR
6. Set the FUNCTION
switch to FREQ A/B.
AUTO
1000000.0
us
Set the TIME
BASE
switch as shown below
lus
1000000.0
ES
and observe the display.
lus
00000010
us
10 us
000.00100
-
s
TIME BASE
DISPLAY
ANNUNCIATOR
1 ms
0000.1000
8
ims
00010.000
$
AUTO
1000000.0
u
10 ms
001000.00
s
1
10000000
‚is
0100000.0
S
10
0000001.0
is
10000000
S
102
000001.00
10s
e000000.00
ks
108
00001 .000
104
0001.0000
e = Overflow
10°
001.00000
108
01000000
Uu
Time Interval Self Check
sn
"i
S
9. Set the FUNCTION switch to TIME A-B.
Set the TIME
BASE
switch
as shown
below
s= Ovidio
and observe the display.
TIME BASE
DISPLAY
ANNUNCIATOR
AUTO
1000000.5
MS
A ps
1000000.5
us
l us
00000015
us
i
|
10 us
000.00105
s
Period Average Self Check
Ta
0000.1005
>
7. Set the FUNCTION switch to PER AVG B.
Da
GE
e
Set the TIME
BASE
switch as shown below
10 ms
001000.05
E
and observe the display.
at bo
See
j
TIME BASE
DISPLAY
ANNUNCIATOR
SC
RE
Ke
AUTO
100.00000
Mus
e = Overflow
ei
nu.
i
Time interval Average Self Check
10?
00000.100
Us
10. Set the FUNCTION
switch to TIME AVG
105
0000.1000
Jus
A-B. Set the TIME BASE switch as shown
10°
000.10000
MS
below and observe the display.
105
00100.000
Mus
105
0100.0000
Mus
TIME BASE
DISPLAY
ANNUNCIATOR
10°
100.00000
Mus
108
200000.000
AUTO
100.00000
Mus
1
1000000.0
HS
e = Overflow
10
000000.10
us
102
-00000.100
HS
Mus = Nanoseconds
103
0000.1000
us
104
000,10000
MS
10°
00100.000
Mus
106
0100.0000
Mus
107
100.00000
Mus
Period Self Check
103
*00000.000
8.
Set the FUNCTION
switch to PER B. Set the
e = Overflow
TIME
BASE
switch
as
shown
below
and
observe the display.
Mus = Nanoseconds
Figure 9H-3-4.
Making Self Check Measurements (Cont'd)
9H-3-8

Advertisement

Table of Contents
loading

Table of Contents