2.3 Room Layout Considerations

2.3.1 Service Access

Allow appropriate space for service access of equipment. Consult component pre–installation
directions for clearance information.
2.3.2 Clinical Access
Make sure that you plan the room with the following clinical access requirements:
Provide easy access to the patient table. Stretchers and other mobile hospital equipment
must reach the table quickly.
Gantries installation shall make a provision so that the clearance is 500 mm (19.7in ) around
the Frontal and Lateral Gantries.
The layout of the table in the room shall make a provision so that clearance between
maximum table position (head side) and any object in the room (e.g. wall, device) be
superior to 500 mm (19.68 in) (650 mm (25.5 in) if Head Extender is used).
Provide sufficient space around the patient table for the unimpeded conduct of CPR
(Cardiac Pulmonary Resuscitation). With the table in this position, the table must be capable
of rotating +/- 45°.
Clinicians at the patient table must be able to communicate with assistants in the control
area.
There must be an unrestricted view of the video monitors and physiological monitoring
equipment from the vascular table.
Operators in the control area must have easy access to the control console. However,
position the controls (including handswitches) so that the operator cannot take exposures
while looking around or standing outside the control booth's lead glass window.
Operators in the control area must have easy access to video recorders and injector
programmers, film and video storage cabinets, and service and operating manuals.
Consult customer on the number and location of nonelectrical lines (air, oxygen, vacuum,
water, etc.) in the vascular room.
For Large Display systems, in case of failure of the main monitor, the clearance around the
main monitor suspension must assure that it can be immediately be flipped at 180°,
exposing the backup monitors.
2.3.3 Peripheral Equipment
Consult hospital personnel regarding additional space requirements for the following types of
hospital equipment:
Storage cabinets.
Sinks.
Oxygen stations.
IV apparatus.
Chapter 2 Equipment Requirements
Innova
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
TM
Direction 5694389-1-1EN, Revision 2
103