Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Illustration 3-5: Table floor mounting layout
For alternative table bolts or seismic area, refer to template drawing shown in Illustration
LC Positioner And Table Floor Mounting Template contained in
Requirements.
3.3.4 Hole dimension and preferred location in concrete floor
In the examination room, the Innova LC Positioner is not placed on a computer floor but directly
put on concrete floor, the location of the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to come through
the floor and in this case you will need two holes, one for the LC Positioner and the other for the
patient table.
The diameter of both holes must be the same 225 mm or 9 in.
Chapter 3 Physical Requirements of Room
Prepare the floor such that the Innova IQ Table base plate will be flush with the floor
finish surface, taking into account the thickness of the floor finish material.
Direction 5160944-13-1EN, Revision 2
Chapter 5, Section 2, Mounting
Innova
69