2 Environmental Requirements/Limitations

2.1 Room Climate

Innova system room climate requirements – relative humidity and temperature (individual products
or components are classed by their installation area)
Table 3-1:
INSTALLATION
ROOM OF
PRODUCT OR
COMPONENT
Examination room 30%
Technical room
(See Note (1))
Technical room
with Fluoro UPS
optional
Control room
Innova Digital De‐
tector
NOTE:
(1): The target temperature (best recommended) is 18°C (64°F).
NOTE:
(2): The detector should be stored at 10 to 40 °C (50 to 104 °F) and less than or equal
to 90% RH in the plastic wrapped shipping box. (This should include two bags of
desiccant as well). The lowest temperature (e.g. 10 ºC (50 °F)) and humidity is
preferable. If they are to be stored outside of their shipping box or in the inner shipping
box without plastic wrapping they should be stored at 20 °C (68 °F) or less and 30%
RH or less. In terms of transportation, do not expose to temperatures below –20 ºC (–
4 °F) in its shipping box for more than 15 hours. The detector will reach the ambient
temperature after 20 to 25 hours. The detector should not be allowed to reach
temperatures less than –10 ºC (14 °F) or irreparable damage to the detectors scintillator
will occur. Care must be taken when removing a detector from a shipping box. If the
detector has been subject to cold temperatures for an extended period the detector in
the box should be allowed to sit in the plastic wrapped box to reach room temperature.
This will prevent condensation from occurring. Condensation on the detector can cause
irreparable damage to the electronics. Storage 10 to 40 ºC (50 to 104 °F); 10 to 90 %
RH, 250 day storage transportation –20 to +60 ºC (–4 to 140 °F) and 10 to 80% RH.
The Detector chiller HEAT DRY 1 is shipped within GEMS packaging.
54
Avoid extremes in temperatures
RELATIVE HUMIDITY (NON- CONDENS‐
ING)
IN-USE
MIN
MAX
MIN
70%
10%
30%
75%
10%
30%
75%
10%
30%
75%
10%
See Note (2)
STORAGE
IN-USE (See Note (3))
MAX
MIN
+15°C
70%
+59°F
+10°C
70%
+50°F
+20°C
70%
+68 °F
+10°C
70%
+50°F
Direction 5160944-13-1EN, Revision 2
TEMPERATURE
RECOMMENDED
(See Note (3))
MAX
MIN
MAX
+32°C
Design for Patient/
+90°F
Operator Comfort
+35°C
+13°C
+25°C
+95°F
+55°F
+77°F
+25°C
+20°C
+25°C
+77°F
+68 °F
+77°F
+35°C
+20°C
+25°C
+95°F
+68°F
+77°F
2 Environmental Requirements/Limitations
STORAGE / TRANS‐
PORT
MIN
MAX
-10°C
+50°C
+14°F
+122°F
-14°C
+50°C
+8°F
+122°F
-5°C
+50°C
+23°F
+122°F
-10°C
+50°C
+14°F
+122°F