Chapter 4. Environmental Requirements

4.1 Relative Humidity and Temperature

NOTICE
Avoid extremes in temperatures
Innova system room climate requirements – relative humidity and temperature. Individual products or
components are classed by their installation area for in use see
Temperature on page
Individual products or components are classed separately for Storage, see
Handling Requirements on page
NOTE
Due to the differential in temperature / humidity between the clinical room and the room
below, the floor creates condensation within the gantry / patient table resulting in part
failure or rusting inside the gantry / patient table.
For systems that are planned to be installed on higher level (second floor or above), the
temperature and humidity of the rooms that are directly below the gantry room should be
maintained same as the exam room requirement. Failure to do so will void the equipment
warranty. Avoid above grade installations where the temperature is high in the area below
the gantry / table cables exit if at all possible.
Table 4-1 In use - Humidity and Temperature
INSTALLA-
RELATIVE HUMIDITY (NON-
TION ROOM
OF PRODUCT
OR COMPO-
NENT
MIN
Examination
30%
room
Technical
room (See
30%
Note (1))
Technical
room with
30%
Fluoro UPS
optional
Control room
30%
5750182-1EN 3
129,
25.
CONDENSING)
IN-USE
MAX
70%
75%
75%
75%
TM
Innova
IGS 6
Table 4-1 In use - Humidity and
IN-USE (See Note (2a))
MIN
MAX
+15°C +59°F
+32°C +90°F
+10°C +50°F
+32°C +90°F
+20°C +68 °F
+25°C +77°F
+15°C +59°F
+35°C +95°F
Environmental Requirements
1.4 Product Storage and
TEMPERATURE
RECOMMENDED (See Note
(2b))
MIN
Design for Patient/ Operator
Comfort
+13°C +55°F
+20°C +68 °F
+20°C +68°F
MAX
+25°C +77°F
+25°C +77°F
+25°C +77°F
129