Download Print this page

Specifications - Kenwood FG-273 Service Manual

Function generator
Hide thumbs Also See for FG-273:

Advertisement

SPECIFICATIONS
Frequency Characteristics
Outputs
Sine, square, triangle, pulse, ramp, TTL/CMOS square wave
Frequency range
0.02 Hz to 2 MHz in 7 frequency ranges ( 1/10/100/1 k/1 Ok/1 OOk/1 M)
Accuracy ( 1 )
±
1 digit, 4-digit max. (digital readout to output frequency)
Accuracy (2)
±
5% of full scale (0.2 Hz to 2 MHz) (frequency dial to output frequency)
External frequency control (VCF)
Input voltage
0 to + 10 V DC. Frequency decreases with positive voltage
Variable frequency range
Greater than 1000:1
Variable symmetry
Variable over 1:1 to 40:1 range
DC offset
Continuously variable, maximum of
±
10 V open circuit,
±
5 V into 50 ohms.
Polarity
Inverted or non-inverted
Sine Wave
Distortion
Less than 1%, 10 Hz to 100 kHz
Amplitude flatness
Within
±
1 .0 dB to 1 00 kHz at maximum output amplitude
Output
Variable amplitude
Square Wave
Symmetry
Less than
±
3% at 100 Hz
Rise and fall time
Less than 1 00 ns at maximum output
Output
Variable amplitude
Triangle Wave
Linearity
Less than 1 % at 1 00 Hz
Output
Variable amplitude
TTL Output
Rise and fall time
Less than 25 ns
Output
TTL level
CMOS Output
Rise and fall time
Less than 60 ns
Output
+5 to + 1 5 V, continuously variable
Sweep Characteristics
Internal
Linear or logarithm
Sweep rate
0.5 Hz (2 s) to 50 Hz (20 ms), continuously variable
Sweep width
Variable from 10:1 to 1000:1
External sweep
Front panel VCF jack, Input impedance is 11 . 5 kO.
3

Advertisement

loading