HP 5315 A Operating And Service Manual page 86

100 mhz universal counter
Table of Contents

Advertisement

Model
5315A/
B
Serv ice
8-'4
8-83.
Test Procedure 3, Internal Reference Oscillator
6·84.
To
verify
the internal reference osci llator
check
for
the
presence of the 10 MHz
sin
ewave
at A1U2(21
); see the
following figure for typ
ica
l
waveform:
20
J,l.sec
If
the
10
MHz
reference
oscillator signal
is
not
present, check A1Y1,
Q2, and Q1
.
8-85. Test Procedure 4,
Input
Amplifiers
8·86.
To
verify
the
proper
operation of A3/A4
Input
Amp
lif ier,
apply
a
10
M
Hz
sine wave at
25 mV
rms
("70
mV
pop)
to
INPUT
A w ith
5315A/
B in
COM A.
Referring
to
A3/
A4
schemati
c
diagram,
Figure
8-8,
check
that
the
proper
waveforms exist at
each
channel output
(Test Points
E
and J)
per the following
examples.
If
either
is
not
present, use
the examp les to
trace
the
signal
back.
TEST
POINT
A
Cha
nnel
A
Signal
Conditioning
Output
A4Q4(7)
B
Channel
A ComparalOr
Input
A4U2{7)
C Channel A Schmit! Trigger Output A4U2(1
)
D
Channel
A
Trigger
Amplifier
One-Shot
Input
A4U3(2)
E Channel
A to
MRC
A4
P2(S)
F
Channel 8
Signal Conditioning
O utput
A4Q2(7)
G Channel B
Comparator
Input
A4U2(10)
H Channel
8
Schmitt
Trigger
Output
A4U2(16)
I
Channel 8
Trigger Amplifier
One-Shot Input
A4U3(6)
J
Channel
8
to MRC
A4P2(6)