Hitachi V-212 Operation Manual page 20

Hide thumbs Also See for V-212:
Table of Contents

Advertisement

If the instrument is not used with the power supply turned
on rotate the INTENsity counterclockwise to reduce the
brightness and also blur the FOCUS.
NOTE
For usual observation, leave the following non-calibrating
function section set to "GAL" position.
VARIABLE
Rotate in the direction of arrow.
In this case the VOLTS/DIV is calibrated to
its indicating value.
SWF VAR
Leave the knob in depressed state.
In this case the TIME/DIV is calibrated to
its indicating value.
Align the base line with the horizontal scale line at the
center of the screen by operating CHI POSITION. In some
cases the base line may be oblique to the scale slightly by the
effect of earth magnetism. In this case, bring the base line
until it Ues on the horizontal scale line at the center of the
screen by properly adjusting the semi-fixed variable resistor
TRACE ROTATION on the front panel.
GENERAL MEASUREMENT
(1) In the case of observing a single waveform.
Use CHI or CH2 when not observing the phase difference
between two waveforms or when engaging in a operation
other than X-Y operation. Make the following settings
when using CHI.
MODE select switch
CHI
Trigger MODE switch @
AUTO
Trigger SOURCE (25)
INT
INT TRIG (V-212 only) (g)
CH1
Under these settings almost all the repetitive signals of
about 25 Hz or more applied to CH 1 can be synchro
nized and observed by adjusting trigger LEVEL. Since the
MODE of horizontal axis is at AUTO position, the base line
appears even when no signal is present or when input
coupling switch is at GND position. This means that the DC
voltage can be measured. The following switching is needed
when observing low frequency signals of about 25 Hz or less.
Trigger MODE
NORM
Synchronization can be effected by operating LEVER knob
under this setting.
16

Advertisement

Table of Contents
loading

This manual is also suitable for:

V-211

Table of Contents