Electrical Representation Of Measuring Results (Optional); Remote Control; Warm-Up Time; Calibration - Philips PM2528 Operating Manual

Automatic rms multimeter
Hide thumbs Also See for PM2528:
Table of Contents

Advertisement

20
2.4.3.
Electrical representation of measuring results
Digital output (optional)
Output system
1EC-Bus interface PM9291
Parallel BCD output PM9292
Output
Galvanically separated from input
Maximum
voltage between
250Vrms, 350Vpeak
0 terminal and output
For detailed specifications of the PM9291 and PM9292 refer to section 3.7 {page 31) and section 3.5 (page 28)
respectively,
Analog output (optional)
Output system
Analog output PM9254, digital to analog converter
Output
Galvanically separated from the input
Output voltage
o-1V
Maximum voltage between
250Vrms, 360Vpeak
Q terminal and output
For detailed specifications of the PM9254 refer to section 3.6 (page 30}.
2.4.4.
Remote control
Via 1EC-bus interface PM9291
Alt functions, ranges and modes can be controlled via the PM9294
(optional)
Via BCD paralle) output PM9282
A start command to start a measurement can be given via the
(optional)
P9292
Via BNC at rear side of PM2528
A start command to start a measurement can be given via the BNC
(START EXT) connector
For detailed information on the PM9291 and PM9292 refer to section 3.7 (page 31) and section 3.6 (page 28)
respectively,
2.4.5,
Warm-up time
Warm-up time
30 minutes to reach specified accuracy
2 hours before calibration
2.4.6.
Calibration
Recalibration interval
180 days