Wrong Setting Information Indicating; Wrong Connection Information Indicating; Error Of Components And Solving Method - Epson Weighing Indicator XK3190-A1+P User Manual

Epson weighing indicator user's guide
Table of Contents

Advertisement

6.3

Wrong setting information indicating

1 Err 13
Wrong setting of the division value .
2 Err 14
Decimal point must be less than 5, please reset the decimal bit.
3 Err 15
Overload warning must be > 100, please reset it.
Err 17
Enter new data which should be less than 65000.

6.4 Wrong connection information indicating

1 Err
P
It means the printer has trouble or is wrongly connected. push any
key to quit.
2 Err 01
It means the load cell signal line is wrongly connected, or its signal is
negative.
(1) If this scale is under usage , then can be sure : the load cell connecting wires
had troubles, or load cell has been damaged.
(2) If this scale hasn't been calibrated, the user should check the load cell's
connective wires first. ( Whether the signal line has been negatively
connected). If the loadcell cable is corretly linked, but the indicator still
indicates Err01 , the problem can then be solved by short connecting pin 4
and pin 6.
3 Err 02
It means the load cell's wires are wrongly linked, or the signal value
exceeds the A/D converting range.
(1) If this scale is under usage, then can be sure: the load cell's connective wires
had troubles, or load cell is damaged.
(2) If this scale hasn't been calibrated, please check as following:
a) Carefully check the load cell's connective wires is right or not.
b) Check if load cell is suitable or not: It should satisfy the following terms: the
"scale's dead load + scale's rated capacity " must be less than the load cell's
rated capacity.
c) This may happen in case that there are more than one load cell in the weighing
system.
d) Short connecting the pin 4 and pin 1 on the socket of the load cell.

6.5 Error of components and solving method

1 Err 18
Key board has problems, It will indicates for ten seconds, then
indicator enters weighing mode.
2 Err 20
The data is partly lost in RAM. Operator should put in the calibration
plug to self-check, and then pull it out.
X K 3 1 9 0 - A 1
+P