TEMPERATURE SETTING

Chamber temperature
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
o
temperature is -35
C.
Note: The unit is set at the factory that the chamber temperature -30
Table 1. Basic operation sequence (Example: Chamber temperature -35
Description of operation
1
Switch ON the freezer.
2
Press SET key.
3
Press
key.
Press
key and scroll the figure
4
to 5.
5
Press SET key.
Note:
The temperature set mode returns to the temperature display mode automatically when 90 seconds has
passed without any key operation.
Although the value of the chamber temperature setting can range between -18
guaranteed temperature without load is -35
Key lock function
This unit is provided with the key lock function. When the key lock is ON, change of temperature setting
through the key pad is not available. The key lock is set in OFF at the factory.
Display
Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation
1
2
Press
key for 5 seconds.
Press
key and scroll the figure
4
to 1.
5
Press SET key.
Key operated
o
C at ambient temperature of 35
Mode
Key lock is OFF
Key lock is ON
Key operated
o
C.
Indication after operation
The current chamber temperature is
----
displayed.
The second digit is flashed.
SET
When pressed, the settable
digit moves.
When pressed, the figure of settable
digit increases.
Set temperature is memorized and the
SET
current chamber temperature is
displayed.
Function
Enable to change of temperature setting
Disable to change of temperature setting
Indication after operation
----
The current chamber temperature is
displayed.
The right digit is flashed.
When pressed, the figure of settable
digit increases.
The key lock is set to ON.
SET
The current chamber temperature is
displayed.
16
o
C)
o
o
C and -40
C, the
o
C.