Alarms & Safety Functions - Sanyo MDF-136 Instruction Manual

Biomedical freezer
Hide thumbs Also See for MDF-136:
Table of Contents

Advertisement

ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions.
Table 5 Alarms and safety functions
Alarm & Safety
If the chamber temperature is higher
High temperature
than the temperature at which the
alarm
high temperature alarm is activated.
If the chamber temperature is lower
Low temperature
than the temperature at which the
alarm
low temperature alarm is activated.
In the case of power failure.
When power switch is turned OFF.
Power failure alarm
When the power to the unit is
disconnected.
When there is no key pressing in
Auto-return
each setting mode for 90 seconds.
Key lock
When the key lock is "ON".
If the thermal sensor is disconnected.
Thermal sensor
abnormality
If the thermal sensor is short-circuited.
If the protective sensor for compressor
Thermal sensor
is disconnected.
abnormality
(MDF-436 only)
If the protective sensor for compressor
is short-circuited.
Battery switch
When the battery switch is OFF at the
check
time of alarm test.
In the case of failure of compressor
cooling fan motor.
Compressor temp.
In the case of abnormal high
abnormality
temperature due to the dust on the
(MDF-436 only)
condenser.
In the case of abnormal high ambient
temperature.
Note:
1. The above power failure alarm is available when the battery switch is ON and the battery is charged.
If the battery switch is OFF or the battery is discharged, only the remote alarm is activated.
2. The power failure alarm can be kept about 12 hours with the battery charged completely. 2-day
operation of the freezer is needed to charge the battery full.
3. The chamber temperature is displayed for 5 seconds if the BUZZER key is depressed during the power
failure alarm. After that, the alarm stops and the alarm lamp blinks.
4. The remote alarm is silenced by pressing BUZZER key as the remote alarm is operated in conjunction
with the buzzer, except for the power failure alarm.
5. After power failure, the operation is resumed with the condition before power failure since the
temperature setting and alarm temperature setting are memorized in a nonvolatile memory.
Situation
ALARM lamp is flashed.
Temperature indicator is
flashed.
ALARM lamp is flashed.
Temperature indicator is
flashed.
ALARM lamp is flashed.
Chamber temperature is
displayed.
ALARM lamp is flashed.
E01 and chamber temp. are
displayed alternately.
ALARM lamp is flashed.
E02 and chamber temp. are
displayed alternately.
ALARM lamp is flashed.
E03 and chamber temp. are
displayed alternately.
ALARM lamp is flashed.
E04 and chamber temp. are
displayed alternately.
E09 is flashed.
E10 is flashed.
18
Indication
Buzzer
Intermittent tone with
15 minutes delay.
Intermittent tone with
15 minutes delay.
Intermittent tone
-----
-----
-----
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Intermittent tone
Safety operation
Remote alarm with 15
minutes delay.
Remote alarm with 15
minutes delay.
Remote alarm.
Finishing of each
Setting mode.
Change of setting is
disable.
Remote alarm.
Continuous running.
Remote alarm.
Continuous running.
Remote alarm.
Normal operation.
Remote alarm.
Normal operation.
Remote alarm.
Remote alarm.
Compressor stops
running.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdf-436Mdf-236

Table of Contents