Equipment Requirements; Equipment Required - HP 3325A Calibration Manual

Synthesizer/function generator
Table of Contents

Advertisement

TB 9-6625-2049-35
Test instrument parameters
Amplitude
accuracy
offset
and
no
(range 1)
into 50
Attenuator accuracy (these
errors are additive with the
amplitude accuracy errors)
Amplitude output
Amplitude (option 002)
(high voltage output)
Dc offset
Dc offset (option 002)
Dc plus ac
1
Not verified below 50 Hz.
2
Except lowest attenuator range where accuracy is 20 V.
4. Equipment Required. Table 2. identifies the specific equipment to be used in this
calibration procedure.
Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity. The
item selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.
4
Table 1. Calibration Description - Continued
with
dc
attenuation
load
Sine wave: .001 Hz to 100 kHz
Square: .001 Hz to 100 kHz
Triangle: .001 Hz to 2 kHz
2 kHz to 10 kHz
Ramps: .001 Hz to 500 Hz
500 Hz to 10 kHz
Function and frequency range
.001 Hz to 100 kHz
Attenuator ranges 2 through 8
100 kHz to 10 MHz
Attenuator ranges 2 through 8
10 MHz to 20 MHz
Attenuator ranges 2 through 4
Attenuator ranges 5 through 8
Range: 1.000 mV to 10.00 V p-p
Range: 4 mV to 40 V p-p (>500 )
Accuracy: +2% of full output for each range at 2 kHz
Flatness: +10% of programmed amplitude
Range: +5 V dc
Accuracy: +0.4% of full peak output for each attenuator range
Range: +20 V dc
Accuracy: + (1% +25 mV) of full output for each attenuator range
Range: <1 MHz
Accuracy: +1.2%
Ramps: +2.4%
Range: >1 MHz
Accuracy: +3%
SECTION II

EQUIPMENT REQUIREMENTS

This equipment is issued with Secondary Transfer Calibration
Performance specifications
Tolerance relative to
programmed amplitude
0.3 dB
3%
4%
6%
4%
11%
Tolerance relative to
programmed amplitude
0.1 dB
0.2 dB
0.5 dB
0.2 dB
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

3325b