Amplitude Modulation Distortion; Square Wave Characteristics - HP 3325A Calibration Manual

Synthesizer/function generator
Table of Contents

Advertisement

9. Amplitude Modulation Distortion

a. Performance Check
(1) Press keys and enter values using DATA keys as listed in (a) through (i) below:
(a) SIGNAL off (option 002 not lit).
(b) FUNCTION sine wave.
(c) ENTRY FREQ.
(d) 1 MHz.
(e) ENTRY AMPTD.
(f) 3 VOLT.
(g) ENTRY DC OFFSET.
(h) 0 VOLT.
(i) Blue then AM ON.
(2) Connect equipment as shown in figure 2.
(3) Set audio analyzer for a 50 (SPCL 75), 10 kHz, 2.5 V output.
(4) Set measuring receiver to measure AM.
(5) Adjust audio analyzer level until measuring receiver indicates 80%.
(6) Set audio analyzer to measure distortion in dB. Audio analyzer indication will
be -30 dB.
(7) Press TI blue key then DATA AM OFF key.
b. Adjustments. No adjustments can be made.

10. Square Wave Characteristics

a. Performance Check
(1) Press keys and enter values using DATA keys as listed in (a) through (f) below:
Figure 2. AM distortion - equipment setup.
TB 9-6625-2049-35
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

3325b

Table of Contents