Model 2400KL Thank you for your purchase of this Health o meter® Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Caution and Warning....................3 Specifications for This Scale ..................5 Certifications / Connectivity / Disposal ..............6 Assembly Instructions 2400KL Scale ...............7...
CAUTION AND WARNING To prevent injury and damage to your scale, please follow these instructions very carefully. Assemble and operate the scale per the enclosed user instructions. Operate this scale exclusively with the AC adapter provided by Health o meter.
Page 4
CAUTION AND WARNING To prevent patient injury, the patient must be attended throughout the entire weighing event. Exercise caution when rolling patient onto and off of the scale. To prevent injury, exercise care when adjusting wheelchair ramps. To prevent injury and damage to your scale, do not drop or mishandle scale during handling or transport.
SPECIFICATIONS FOR THIS SCALE General The Model 2400KL, portable and adjustable patient scale uses highly sophisticated microprocessor technology. Each precision scale is designed to provide accurate, reliable, and repeatable weight measurements. In addition, each scale is designed to provide the user with features that make the weighing process simple, fast, and convenient.
Support, available during normal business hours at (+1) 800-815-6615 or (+1) 708-377-0600. 2400KL Scale Disposal This Health o meter® Professional scale must be disposed of properly as electronic waste. Follow the national, regional or local regulations which apply to you for disposal of electronic...
2400KL ASSEMBLY INSTRUCTIONS Before Assembly Each 2400KL scale is shipped disassembled in a single carton. Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact your shipper or a Health o meter® Professional representative immediately. Claims must be filed with the shipper as soon as possible after receipt of the package.
Page 8
2400KL ASSEMBLY INSTRUCTIONS (CONTINUED) Assembling the 2400KL Scale (Refer to Figures 1-7) Caution: It is recommended that the assembly be performed by two persons. 1. Place scale assembly on a flat, level, and dry surface, and carefully turn the ramp upside down.
DISPLAY ASSEMBLY WALL MOUNT INSTRUCTIONS Display Assembly Wall Mounting Instructions Caution: When wall mounting the display assembly, ensure that the mounting location is free of electrical lines, gas pipes, and other safety hazards. Ensure that wall is of construction that can support the display assembly.
2. Remove the battery cover on the side of the display assembly by unscrewing the (2) battery cover screws (see Figure 11). 3. Replace or install the batteries with new units. Health o meter recommends using an alkaline or lithium battery, rated for 130°F (54°C) operation. Do not use zinc-carbon batteries.
Page 15
(see Figures 8 & 9). 5. For the 2400KL, grasp the handle, located on the center of the ramp, and lift the scale into the upright position. The scale can now be transported (see Figure 10).
QUICK START / OPERATING INSTRUCTIONS Keypad Figure 12 Function Description ON/OFF Turns scale on and off. ZERO Zeros the scale prior to a weighing. Holds the weight value of the patient on the display until the button is HOLD/RELEASE pressed again to clear the value. Also moves the cursor from left to right ►...
Page 18
The Tare Function automatically performs this subtraction. The 2400KL scale allow tare weight to be entered manually by using the keypad, or automatically, as outlined in this section.
Page 19
Exercise caution when rolling patient onto and off of the scale. 5. Once a weight has been displayed, while the patient is still on the scale, Health o meter recommends performing a second weighing by pressing the REWEIGH button.
Page 20
5. Once a weight has been displayed, while the patient is still on the scale, Health o meter recommends performing a second weighing by pressing the REWEIGH button.
Maintenance The following pages provide instructions for maintenance, cleaning, calibrating, and troubleshooting for the 2400KL scale. Maintenance operations other than those described in this manual should be performed by qualified service personnel. Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function.
This scale has been factory calibrated, and does not require calibration prior to use. For maintenance, the scale can be calibrated. Calibration of your 2400KL scale is performed using kilograms (kg) or pounds (lb), according to the units of measure used upon entering into the calibration path.
TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and to correct any failures. Symptom Possible Cause Corrective Action Scale does not turn on 1. Dead battery 1. Replace batteries 2. Faulty electrical outlet 2. Use a different outlet 3.
What does the Warranty Cover? Health o meter® Professional scales are warranted from date of purchase against defects of materials or in workmanship for a period of two (2) years. If product fails to function properly, return the product, freight prepaid and properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for instructions).