Model 2600KL / 2610KL Thank you for your purchase of this Health o meter® Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Caution and Warning....................3 Specifications ......................4 Certifications / Connectivity / Disposal ..............5 Assembly Instructions ....................
CAUTION AND WARNING INTENDED USE ® This Health o meter Professional wheelchair scale is intended to be used in a professional medical environment by trained medical staff. This product was designed to weigh patients who are safely seated in a wheelchair or standing on the platform.
Bluetooth wireless technology. This connectivity helps minimize a source of potential medical errors caused by incorrectly copying and then recording patient data. For more information regarding how to connect the scale to other electronic devices, please contact Health o meter® Professional Scales Technical Support, available during normal business hours at 1-800-638- 3722.
ASSEMBLY INSTRUCTIONS Before Assembly Each 2600KL/2610KL scale is shipped disassembled in a single carton. Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact your shipper or a Health o meter® Professional representative immediately at 1-800-815-6615. Claims must be filed with the shipper as soon as possible after receipt of the package.
Page 7
ASSEMBLY INSTRUCTIONS (CONT.) Health o meter® Professional Scales recommends that assembly is performed by two people. Warning: Do not drop scale, as this can permanently damage the load cells. Caution: Do not insert the batteries into the display head until it is mounted onto the scale.
Page 8
ASSEMBLY INSTRUCTIONS (CONT.) 6. Obtain the display head and align the four holes on the back of the display head with the four mounting holes on the silver display support. Using the (4) included M4 x 0.7 x 10mm screws, fasten the display head onto the display support using a Phillips head screwdriver to tighten the screws.
Page 9
Ramp Assembly Instructions Follow these instructions for moving the ramp to the opposite side of the scale, or for adding a second ramp to the scale (2610KL or B2600RAMP option): 1. On the side of the platform opposite to the ramp, remove the guardrail mounted...
AC adapter into the power source. Warning: To avoid a safety hazard, do not use an unspecified AC adapter. Only use the Health o meter® Professional AC adapter model ADPT30 designed for this scale. Power Adapter Jack For battery operation, see Battery Installation below.
Page 11
SET-UP INSTRUCTIONS (CONT) The scale must be powered on before performing any of the following operations. Press to turn on the scale. Unit of Measure Configuration Unit The weight measuring unit (pounds/LB or kilograms/KG) can be changed by pressing To lock the weight measuring unit, follow this procedure. 1.
Page 12
After each use, the scale will clear the weight of the patient and set itself back to the initial ZERO ® level it found on start-up. Health o meter Professional recommends keeping this setting as Enable. Note: Hitting a key to wake the scale from sleep mode will not prompt the scale to zero.
OPERATING INSTRUCTIONS Key definitions can also be found by pressing and selecting “Key Definitions” in the Help submenu. Keypad Function Description Powers scale on and off. Zeroes the scale prior to a weighing. Also used as a screen navigation key to exit a Zero/ display menu.
Page 14
OPERATING INSTRUCTIONS Tare Function When using this scale, the weight of an object, such as a wheelchair, can be subtracted from the total weight in order to determine the weight of the patient alone. The Tare Function automatically performs this subtraction. The scale allows a tared weight to be entered manually by using the keypad, or automatically, as outlined in this section.
Page 15
OPERATING INSTRUCTIONS (CONT) Weighing a Patient and Taring a Known Weight / Pre-Tare Pre-Tare can be performed within the following weight limits: 0.2 lb - 200 lb / 0.1 kg – 91 kg) 1. Wait until “0.0” appears on the display. Tare 2.
Page 16
OPERATING INSTRUCTIONS (CONT) Weighing a Patient Standing on the Platform 1. Make sure there is nothing on the weighing platform and the display shows “0.0”. 2. Ask the patient to step on the scale. The scale will begin to display the patient’s weight and the word “PROCESSING”. Depending on the movement of the patient on the scale, it may take several seconds for the scale to lock onto the weight.
Page 17
OPERATING INSTRUCTIONS (CONT) BMI Function 1. Make sure there is nothing on the weighing platform and the display shows “0.0”. 2. Follow weight measuring procedure. After obtaining the patient’s weight, and with the patient remaining on the scale, press 3. The scale will prompt entry of the patient’s height. Use the arrow keys to enter the patient’s height.
Page 18
OPERATING INSTRUCTIONS (CONT) Parent/Child Function ® The Parent/Child function on Health o meter Professional scales provides the convenience of weighing a small child or infant while being held by a parent. Follow these simple steps to subtract the parent’s weight from the total so only the child’s weight is displayed.
MAINTENANCE Maintenance The following pages provide instructions for maintenance, cleaning, calibration, and troubleshooting for your scale. Maintenance operations other than those described in this manual should be performed by qualified service personnel. Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function.
CALIBRATION This scale has been factory calibrated, and does not require calibration prior to use. Calibration of your scale can be performed using calibrated kilogram (kg) or pound (lb) weight standards. Minimum calibration is 200 lb / 100 kg. Maximum calibration weight is 1000 lb / 454 kg.
Page 21
CALIBRATION (CONT) 7. When the zeroing process is complete the display will prompt for entry of the calibration weight. Minimum calibration is 200 lb / 100 kg. Maximum calibration weight is 1000 lb / 454 kg. Calibration weight can be adjusted in 10 lb increments.
TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and to ® correct any failures. For further assistance, contact Health o meter Professional Scales Technical Support at 1-800-638-3722. Symptom Possible Cause Corrective Action 1. Dead battery 1.
What does the Warranty Cover? These Health o meter® Professional scales are warranted from date of purchase against defects of materials or in workmanship for a period of two (2) years. If product fails to function properly, return the product, freight prepaid and properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for instructions).