8.2 Setup of Limit Values by Actual Quantity Loads
1
Starting the limit function
L. SEt
Continuous pressing
Key released
Press and hold down the | Set | key.
Release the key when [L.
5 E t ] is
displayed.
The currently set lower-limit value
flashes.
2
Loading the sample for the lower-limit value
•4*
\ \ \ I /.
n n n n
.
u. u u u
i
/
/
I
\
\
Load the sample of the lower-limit value
on the measurement pan.
3
Saving the lower-limit value
3.8 3 Sg
y u
Press the | Memory | key.
After the lower-limit value has been
saved, the balance displays it briefly and
proceeds to the following setup,
i&lf One-point setup was chosen, the
setup is complete.
4
The upper-limit value
setup
H. SEt
M
\ \ \
anoo
The display now changes to
[K
SEt],
indicating that the upper-
limit value can be set.
The currently set upper-limit value
flashes.
5
Loading the sample of the upper-limit value
\ \ I / / /
n n n n
U.UUU
g
/
I
I
\
\ s
6
Saving the upper-limit value
Load the sample of the upper-limit value
on the measurement pan.
Press the | Memory | key.
After saving the upper-limit value, the
balance displays it briefly and
terminates the setup.