15. Initialization of non-volatile memory
0
In Top Screen, press MENU key and select "Tools".
1
Press ENTER key to display "Select Tools (1/2)" and press MENU key to select "Svc".
2
Press ENTER key to display "Password" and input service code "384".
3
Press ENTER key to display "Select Tools Svc" and select "Analog Calibration etc".
4
Press MENU key and select "OK".
5
Press ENTER key to display "Analog Calibration etc" and input "925" in the right side of
"Function" column.
6
Press MENU key and select "OK".
7
Press ENTER key to return to "Select Tools Svc" display. Select "Initialize data".
8
Press MENU key and select "OK". Press ENTER key to display "Confirmation".
9
Press MENU key and select "OK" to commence initialization.
Press MENU key and select "Cancel" to cancel initialization.
10
After commencing initialization, a device automatically returns to Top Screen.
After cancelling initialization, a device returns to "Select Tools Svc".
Procedures
- 59 -
MENU key
Tools
Svc
OK
OK
OK
OK
OK
Cancel