Chapter 8: Maintenance
8.
Pull the plunger all the way out of the syringe barrel.
9.
Fold a disposable lint free paper wipe in quarters. Remove the stopper from the lubricant container and
place the folded wipe over the opening of the container.
10. Turn the container over and allow an air bubble to rise to the base of the container. Turn the container
right side up, wipe the opening and replace the stopper. Store the lubricant in an upright position.
11. Swirl the syringe tip once in the slightly moistened wipe and reinstall it into the syringe barrel. Cycle two
or three strokes to ensure adequate distribution of the lubricant.
GE Analytical Instruments ©2013
Figure 51: Pulling Out the Plunger
Figure 52: Removing the Stopper
Figure 53: Distributing the Lubricant
162 of 238
DLM 68100-06 EN Rev. A