INCUBATOR COMPONENTS
1. Outer door: Sticks to frame with magnetic seal. Door heater is installed in the door panel. The
door opening is reversible. Contact Sanyo representative or agent to change the door hinge from left to
right or vice versa.
2. Inner door: Made of tempered glass, however avoid excessive impact on the glass.
3. Leveling foot: Screw type for adjusting the height. Adjust the foot so that the unit can be level.
4. Tray: Can be pulled toward you.
5. Side support: Right and left side supports can be removed for disinfection. See page 33 and 34.
6. Rear duct: Flow path for circulating air. Removable.
7. Fan (inside the rear duct): Made from polypropylene resin. Can be sterilized by an autoclave.
8. Sample air outlet: This also functions as an internal gas outlet. Use only a supplied cap(9).
9. Sample air outlet cap: Always attach this cap except at the time of using of sample air outlet.
10. Connecting port for CO
cylinder changeover accessory) is installed, both A and B are available. If MCO-5GC is not used, only A
is available. Refer to page 21 for gas cylinder connection. Ensure that the gas pressure is set at
0.03MPaG (0.3kgf/cm
11. Connecting port (A/B) for N
Ensure that gas pressure is set at 0.05MPaG (0.5kgf/cm2G, 7.1psiG). Refer to page25 for automatic
cylinder changeover.
12. Door switch: Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
and N
2
optional UV system kit MCO-18UVS2 is installed).
13. Humidifying pan: Use sterile distilled water to fill the pan.
14. Humidifying pan cover: Prevents UV light being exposed to the chamber. When filling the pan,
lift the front side and take out the pan. See page 35 for details.
15. UV lamp (When an optional UV system kit MCO-18UVS2 is installed):
Sanyo UV lamp does not generate ozone. Never look at the UV light directly. For replacement, contact
Sanyo representative or agent.
16. Water level sensor for humidifying pan: Detects the water level in the humidifying pan. See
page 28 for details.
17. Remote alarm terminal: Refer to page 15.
18. Access port: When not in use, cap with the rubber cap on both outside and inside.
19. Power switch: Main switch of the unit. Also functions as an over-current breaker. The switch is
covered by a switch cover to prevent the accidental push. To turn on or off the switch, remove the
switch cover by loosening the screw. See figure on the right.
20. Glow starter (When an optional UV system kit MCO-18UVS2 is installed): For UV lamp (model; FG-7P)
21. Removal Power supply cord
22. Handle: Outer door handle. When moving the handle from upper to a lower position remove it from
original position first, then remove two cups and screws at the lower position and attach the handle there.
23. N
or O
gas injection nozzle
2
2
gas pipe (rear side): When an optional component MCO-5GC (gas
2
2
G, 4.3psiG). Refer to page 25 for automatic cylinder changeover.
/O
gas pipe (rear side): Refer to page 19 for gas cylinder connection.
2
2
/O
when door is open. UV lamp is also deactivated by door opening (When an
2
2
68
12