OPERATING INSTRUACTIONS
3. Following shows the procedure to changeover the gas supply line (N
cylinder) manually.
(Example: Change of CO
Operation
1
Press SET key.
2
Press ENT key.
Press gas supply line
switching key once.
3
Pressing the key again
selects CO
gas cylinder B
2
4
Press ENT key.
Press gas supply line
switching key once.
5
Pressing the key again
selects N
gas cylinder A
2
6
Press ENT key.
7
Press ENT key.
Note:
The procedure 3 above should be skipped when MCO-5GC is not installed.
gas cylinder from B to A, Change of N
2
Key
operated
The left digit on the temperature
SET
indicator blinks.
The left digit on the CO
ENT
indicator blinks.
The blinking CO
lamp changes to continuous ON
and cylinder A is selected.
The setting of CO
is memorized, and the left digit
ENT
on the O
blinks.
The lamp ON of N
supply lamp changes to blinking
and cylinder B is selected.
The setting of N
memorized. HI is displayed on
ENT
the CO
the setting of upper alarm
temperature is displayed.
This is the end of the setting
mode. The current chamber
ENT
temperature and densities are
displayed.
/O
2
gas cylinder from A to B)
2
Indication after operation
density
2
gas supply
2
gas cylinder
2
density indicator
2
/O
gas
2
2
gas cylinder is
2
density indicator and
2
: lamp ON
82
gas cylinder and CO
gas
2
2
Gas supply line
indicator
CO
O
/N
2
2
2
: lamp blinks
26