Dtc Digital Pcb, Version 2 - ABB PGC5000 Series Service Instructions Manual

Process gas chromatographs
Hide thumbs Also See for PGC5000 Series:
Table of Contents

Advertisement

PGC5000
PGC5000 Series Service Instructions

5.3.4 DTC digital pcb, version 2

This DTC digital PCB has eight LEDs (CR1 to CR8) on the upper left-hand edge for basic troubleshooting. CR1 through CR6 are
Green; CR7 and CR8 are Red. Their display condition (On, Off, Solid or Blinking) indicates the board state.
CR1: Green, Blinking
CR2: Green
CR3: Blinking
CR4: Blinking
CR5: Blinking
CR6: Blinking
OFF
CR7: Blinking
Blinking (unison CR1)
CR8: OFF
If the Digital Signal Processor (DSP) system fails during power-up, the LED bank indicates what was being attempted. No green
LEDs (CR1 to CR6) are blinking and CR7 and CR8 are off. CR1 (MSB) through CR6 form a binary code as follows:
000001
CAN initialization error
000010
application initialization error
000011
CAN node/network setup error
000100
no other node on CANBus, still trying
000101
CAN and application timer engaged
000110
reload in progress
000111
CAN disabled, reload jumping to start over
001000
bootloader initialization
001001
bootloader CAN setup
001010
bootloader start hardware setup
001011
bootloader end hardware setup
001100
bootloader received reset application
100001
bootloader flash erase in progress
110011
bootloader flash program in progress
111111
bootloader flash program complete
001110
start board setup
001111
start DSP initialization
010000
end DSP initialization
010001
Analog/Digital initialization
010010
temperature zone setup
010011
temperature conversion setup
010100
number valves detection
010101
end board setup
010111
start DTC initialization
011000
valve initialization
011001
safety switch initialization
011010
PIC initialization
011011
signal testing initialization
011100
purge testing initialization
011101
miscellaneous initialization
011110
end DTC initialization
100000
start setup CANopen
100001
CAN PDO setup
100010
end setup CANopen
100100
solenoid start initialization
100101
solenoid access
100110
solenoid test
100111
solenoid reverse test
101000
solenoid test complete
101001
solenoid end initialization
101111
received reset application
SI/PGC5000-EN, Rev B
Blinks when the downloaded code is running.
Indicates method execution in progress.
RTC signal is detected.
AC Power present.
PIC communication present..
During Initial Power on
Application code running.
T-Rating Error
No power to DTC board
Normal Operation
5 Diagnostics and troubleshooting