PGC5000
PGC5000 Series Service Instructions
INDICATOR
ISSUE
Low-Low
Concentration is below
Concentration
Low-Low Setpoint
Low
Concentration is below
Concentration
Low Setpoint but above
Low-Low Setpoint
High
Concentration exceeds the
Concentration
High Setpoint but below
High-High Setpoint
High-High
Concentration exceeds the
Concentration
High-High Setpoint
RF Low Limit
Calculated response factor
out of range (low) (% of
deviation from previous RF)
RF High Limit
Calculated response factor
out of range (high) % of
deviation from previous RF)
Retention Time
Retention time is below
Low
Low Setpoint
Retention Time
Retention time exceeds
High
High Setjpoint
Validation
Measurement out of range
(% of deviation from known
standard)
Missing
No peak found matching a
Component
component defined in the
analysis
Unknown
Peak detected in the
Component
analysis and not defined in
the method
Sample Flow Lost
Future functionality:
Sample flow is lost
Digin Check TCF
Programmable Input
checks
Psrs Check TCF
Pressure Check failed
Invalid Analysis
Analysis failed validation
Analysis Aborted
1. Analysis does not
complete analysis due to
temperature, pressure
and/or digital input
programmed command.
2. User intervention using
Stop Now command.
3. Related oven issues.
Offline
Schedule Offline: Oven is
powered down
Idle
Schedule Stopped or
paused
SI/PGC5000-EN, Rev B
RESOLUTION (CHECK & CORRECT)
Information Only. Adjust Limits under
the Setup tab as needed.
Information Only. Adjust Limits under
the Setup tab as needed.
Information Only. Adjust Limits under
the Setup tab as needed
Information Only. Adjust Limits under
the Setup tab as needed
Information Only. Adjust Limits under
the Setup tab as needed.
Information Only. Adjust Limits under
the Setup tab as needed
Information Only. Adjust Limits under
the Setup tab as needed.
Information Only. Adjust Limits under
the Setup tab as needed. Possible
leak. Possible defective EPC pressure
sensor. Possible plugged inlet filter.
Information Only. Adjust Limits under
the Setup tab as needed. Flows
incorrectly set. Defective EPC sensor.
Information. Edit analysis as needed.
Check flows. Verify correct operation
operation of all analytical valves.
Verify sample flow to oven.
Restore Sample flow. Replace sample
flow switch as necessary
See Digital inut TCF in Operating
Instructions.
See pressure check TCF in
Operating Instructions.
1. From Analysis tab edit Analysis to
meet requireemnts.
2. User controlled.
1. Check Pressure, Temperature
and/or Digital Inputs. Correct problem
accordingly.
2. User controlled.
3. Check oven related indicators for
more information.
Information Only. Resets when oven
power is restored.
Information Only.
5 Diagnostics and troubleshooting
RESET
SOURCE
When
Component
reported.
When
Component
Reported
When
Component
reported
When
Component
reported
When
Component
recalibrated
When
Component
recalibrated
When
Component
Reported
When
Component
reported
Wen
Component
revalidated
When
Analysis
reported
When
Analysis
Reported
AOC
Analysis
AOC
Analysis
AOC
Analysis
AOC
Analysis
AOC
Analysis
AOC
Schedule
AOC
Schedule
25

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents