attenuation submenu
PROBE CAL tJ.ENU
o t ti!:nuo t i on
t. i me n\J f l
channel
3
4
s t a r t
c o l
c o n t i n ue
abor t
ex i t menu
54�01;W;-Z:?
channel Key
start cal Key
continue Key
abort Key
exit
menu Key
Utility Menu
12-8
The attenuation submenu is used to calibrate channel gain at the probe
can
tip. Channel gain
0.9
attenuation.
•
Below 0.9 the error message Attenuation
displayed. The corrective action
action is
HP5450lA.
is
If
the probe
not connected to the DC CAUBRATOR OUTPUT
the rear panel or the probe attenuation exceeds approximately 250, the
error messageAltenuation
corrective action is to check the connections and recalibrate. If
recalibration is unsuccessful, refer to the
•
If the probe attenuation cali b ration is successful the displayed
message is
Probe Attenuation
into your channel probe setting.
Pressing this key allows you to select a channel to calibrate. By
continually pressing the key you
channels.
When the cbannel you wisb to calibrate has been selected, press start
eal. Tbe advisory will appear at the bottom of the waveform display
area
Connect the DC Col
Pressing this key prompts you for setup requirements.
press continue.
Press this key wben all setup requirements bave been satisfied. The
ocess
actual cah bration pr
the only active front panel key during the calibration process. If
This is
you press this key, the calibration process will be terminated with the
previous calibration factors intact.
Pressing this key will return you to the Utility menu.
be corrected through probe attenuation down to
too high or bad connection
H P 5450lA Service
=
n.nnnnn This
can
increment througb the fields, i.e.
rear
panel bnc to the probe
will begin.
less than 1, see manual for
to recalibrate the
is
is
displayed. The
manual.
value
has been entered
n,
o f
channel
then
on