5.2

Setting initial parameters with LCD interface

Once you powered on the LLT100, you can set the basic
parameters from the Easy Setup configuration menu. This is
one reason why the Advanced mode is strongly recommended:
it allows for configuration of every function in the instrument.
5.2.1 Setting level or ullage parameters
1. Once the LLT100 is on, access Easy Setup in the
configuration menus and press Select.
2. Select the interface language and press Next. English is
selected by default.
3. Select the measurement mode and press Next.
Standard
is
selected
measurement mode based on the intended application of
the instrument being configured (for more information, see
Table 2 on page 17).
4. Select the level unit and press Next. This is the unit used
in level and ullage measurements, and when calculating
volumes.
5. Edit the vessel height (URP, or "ullage reference point") so
that it corresponds to the height of the vessel to be monitored
(for more information, see "General model" on page 15).
6. Edit the sensor offset (for more information, see "Setting
the sensor offset" on page 21).
7. Select the output type for the 4 – 20 mA process value (PV).
8. Edit the Vessel Empty (LRV, or lower reference value) level
(4 mA).
9. Edit the Vessel Full (URV, or upper reference value) level
(20  mA). The default value is the vessel height entered
previously.
10. (if necessary) Enable the Filling Rate feature (for more
information on the filling rate feature, see section  7.4
"Configuring the filling rate" on page 24).
11. (if necessary) Create a tag for the instrument. This "tag",
comprised of alphanumeric characters, is intended to
simplify identifying the location of the LLT100 (on the
network, in the plant, etc.)
12. Once all Easy Setup parameters have been set, press
Next to exit the Easy Setup menu and go back to the
default Operator page.
16 User Guide | OI-LLT100-EN Rev. D
by
default.
You
select
5.2.2 Setting volume parameters
1. Once the LLT100 is on, access Easy Setup in the
configuration menus and press Select.
2. Select the interface language and press Next. English is
selected by default.
3. Select the measurement mode and press Next.
is
Standard
measurement mode based on the intended application of
the instrument being configured (for more information, see
Table 2 on page 17).
4. Select the level unit and press Next. This is the unit used
in level and ullage measurements, and when calculating
volumes.
5. Edit the vessel height (URP, or "ullage reference point") so
that it corresponds to the height of the vessel to be monitored
(for more information, see "General model" on page 15).
6. Edit the sensor offset (for more information, see "Setting
the sensor offset" on page 21).
the
7. Select Volume for the 4 – 20 mA process value (PV) output
type .
8. Select the unit for the Volume output. The liter (l) is the
default value, but other volume units are available.
9. Edit the minimum volume. The default value is 0.
10. Edit the maximum volume. The default value is 1 000 000
liters (or the equivalent in the unit of your choice).
11. (if necessary) Enable the Filling Rate feature (for more
information on the filling rate feature, see section  7.4
"Configuring the filling rate" on page 24).
12. (if necessary) Create a tag for the instrument. This "tag",
comprised of alphanumeric characters, is intended to
simplify identifying the location of the LLT100 (on the
network, in the plant, etc.)
13. Once you have configured the initial volume measurement
parameters, you still need to setup your linearization table
in the Device Setup menu (8 "Configuring linearization" on
page  25) as requested at the end of the Easy Setup
menu.
14. Once all Easy Setup parameters have been set, press
Next to exit the Easy Setup menu and go back to the
default Operator page.