Sanyo MCO-5M Service Manual page 15

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

2.
Temperature control
Setting range
Display range
Setting method
Out of the range:
Control
Alarm
3.
CO
control
2
Setting range
Display range
Setting method
Out of the range:
Control
Control OFF
Alarm
Warming-up
4.
O
control
2
Setting range
Display range
Setting method
Out of the range:
Control
Control OFF
Alarm
Warming-up
:
0 C
50 C
:
0 C
99.9 C
: Press the SET key. Change to the desired value with
key, then press the ENT key. The value will be stored and then the
device shifts to CO
When a value is set out of the range and ENT key is pressed, buzzer
sounds 1 second (continuously) and the previous value is still remained.
: PID control
:
When the PV is SV 1 C or higher, the display digit blinks.
After 15minutes later, buzzer sounds and remote output turns ON.
:
0%~20%
Note) When O
setting value is 70.1% or more, you can only set CO
2
10% or less.
:
0%~99.9%
: Press SET key for twice.
Change to the desired value with
key to store the CO
mode.
When a value is set out of the range and ENT key is pressed, buzzer
sounds 1 second (continuously) and the previous value is still remained.
: PID control
:
When a value is set in 0.0%, CO
Control OFF.
:
When the internal CO
blinks. After the 15mininutes later, buzzer sounds and remote alarm
terminal turns ON.
:
The unit is warming-up for 4 minutes just after the power supplied.
At the time CO
display is shown as "- - -".
2
During warming-up, every control and alarms would be inoperative.
:
1%~18%, 22%~80%
Note) When CO
setting value is 10.1% or more, you can only set O
2
70% or less.
:
0~99%
: Press SET key for three times.
Change to the desired value with
key to store the O
protection temperature setting mode.
When a value is set out of the range and ENT key is pressed, buzzer
sounds 1 second (continuously) and the previous value is still remained.
: PID control
:
When O
value is set in 18.1~21.9%, O
2
be the Control OFF.
:
When the internal O
After the 15mininutes later, buzzer sounds and remote alarm terminal
turns ON.
:
The unit is warming-up for 4 minutes just after the power supplied.
At the time CO
display is shown as "- - -".
2
During warming-up, every control and alarms would be inoperative.
setting mode.
2
key and
value. Then the device shifts to the O
2
display will be disappeared to be the
2
level deviates SV 1% or higher, CO
2
key and
value. Then the device shifts to the overheat
2
2
level deviates SV 1% or higher, O
2
12
key and
key, then press ENT
2
2
key, then press ENT
display will be disappeared to
display blinks.
2
in
2
setting
display
in
2