Sanyo MCO-5M Service Manual page 26

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

<SV and buzzer recovery time>
SV
000
010
020
030
040
050
060
*Optional mode: Remote alarm can be linked with buzzer in optional function mode.
(The function code is not opened to users)
Buzzer setting
Initial: auto recovery
(F25=030)
Inoperative
recovery (F25=000)
Setting
Remote alarm action
000 (initial)
Not linked with buzzer
001
Linked with buzzer
16.
Temperature offset value
The difference between temperature sensed in temperature control sensor and temperature in
internal 1/2H should be offset:
Offset value: +0.8 C
17.
Storing in non-volatile memory
2 times comparison method: When the main power is supplied, a data storing in non-volatile
memory is called for 2 times to compare. If the data 1
called, the data would be stored in RAM. Otherwise, the comparison would be repeated until
the one coincides with another.
18.
Example for displays
Current temperature +36.2℃
Set temperature +37.0℃
Function F03
(Service code input)
Bottom heater ratio 004
Lock Mode
L 0
Buzzer recovery time after buzzer is
cancelled by pressing BZ key
Buzzer does not recover
Approx. 10min
Approx. 20min
Approx. 30min
Approx. 40min
Approx. 50min
Approx. 60min
Buzzer solely activates
(F24=000)
Buzzer: auto recovery
Remote alarm: activating
auto
Buzzer: does not auto recover
Remote alarm: activating
CO
2
CO
2
CO
2
UV lamp additional time
23
Remote alarm action
Buzzer: auto recovery
Remote alarm: auto recovery
Buzzer: does not auto recover
Remote alarm: does not auto
recover
st
called coincides with the data of 2
level +5.0%
level zero adjustment value –1.3%
gas error E01
Linked with buzzer
(F24=001)
nd