60
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Electronics
module").
Caution:
All user-specific settings must be entered again. Hence, you have to
carry out a new setup after the electronics exchange.
If you have stored the data of the parameter adjustment during the
first setup of the sensor, you can transfer these to the replacement
electronics module. A new setup is no more necessary.
9.6
Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
HART modem
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedure in the download area on our homepage:
www.siemens.com/sitranslg.
You can find information about the installation in the download file.
Caution:
Instruments with approvals can be bound to certain software ver-
sions. Therefore make sure that the approval remains effective with a
software update.
You can find detailed information in the download area on our home-
page: www.siemens.com/sitranslg.
9.7
How to proceed if a repair is needed
If it is necessary to repair the instrument, please contact Sie-
mens Milltronics Process Instruments. You find the locations on
"www.siemens.com/sitranslg".
SITRANS LG250 - Operating Instructions
PBD-51041056