ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION
LS4000
Diode laser analyzer
Version for measuring NH
Highest precision under
harshest conditions
Measurement made easy
and H
O
3
2

Advertisement

Table of Contents
loading

Summary of Contents for ABB LS4000

  • Page 1 — ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION LS4000 Diode laser analyzer Version for measuring NH and H Highest precision under harshest conditions Measurement made easy...
  • Page 2: Table Of Contents

    Connecting equipotential bonding ................49   Connecting the power supply ..................50   Gas analyzer start-up ............................. 51   Testing and approving gas analyzer ................. 51   Connecting the supply voltage .................. 53   Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 3 Error messages in "Malfunction" status ..............110   Gas analyzer shutdown ..........................112   Ending operation and shutting down the gas analyzer ........112   Disassembling the gas analyzer ................113   OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 4: Preface

     Data sheets  Technical information  Certificates Internet You will find information on ABB Analytical products and services online at http://www.abb.com/analytical. Service contact If the information in this operating instruction does not cover a particular situation, ABB Service will be pleased to supply additional information as required.
  • Page 5: Safety Instructions

    Applied safety Classification Standard Degree of protection standards Safety of electric devices EN 61010-1 Protection class I Safety of laser devices EN 60825-1 Laser class 1 OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 6: Description

     T/P probe(s) for dynamic temperature and pressure correction (depending on application)  Power supply  Sensors for analog and digital outputs A PC can be temporarily connected to the junction box for service purposes. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 7 Analyzer measuring The LS4000 uses the optical measurement method of tunable diode laser principle absorption spectroscopy (TDLAS), which is based on the fact that gases ab- sorb light of specific wavelengths. In this method, a configurable laser diode in the transmitter unit emits a laser beam, which passes through the process gas and shines onto the photode- tector in the receiver unit.
  • Page 8: Specifications

    AC 100...240 V ± 10 %; 50...60 Hz Power consumption 30 VA IP rating Overvoltage category Degree of pollution Safe isolation Safety extra-low voltage (SELV) on the low voltage side Overload protection Voltage and current limitation Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 9 Safety and Safety Tested to EN 61010-1:2010 electromagnetic Safety in accordance with U.S. The LS4000 gas analyzer is certified for use compatibility and Canadian standards – UL, in “General Purpose” environments. It is in line with the CAN/CSA-C22.2 no. 61010-1-12 and UL no.
  • Page 10: Labels

    The transmitter unit and receiver unit are each fitted with a name plate. receiver unit: deciphering name plates Meaning Details of the manufacturer Model name Transmitter unit, receiver unit Permissible ambient temperature for operation Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 11 Class I infrared laser beam invisible to the human eye. Junction box: interpreting the name plate Meaning Details of the manufacturer Model name F-no. = Manufacturing number, A no. = Order no. Supply voltage and power consumption OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 12 EFUP marking (EFUP = environment friendly use period): 50 years of operating time in accordance with the EU's RoHS Directive without any leaks of substances hazardous to health or the envi- ronment under normal conditions of use CE mark Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 13: Preparing For Installation

    Shut down the system in which the laser analyzer is to be fitted. Wait until the system temperature has fallen to its original level. Ensure that the system is no longer pressurized. Secure the system against startup. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 14: Preparing The Installation Site

    Meaning Dimension Tolerance Outer diameter 152.4 mm Max. ± 0.5 mm Pitch circle diameter 120.6 mm Max. ± 0.3 mm Internal diameter 49.7...65.3 mm Hole diameter 19 mm Max. ± 0.2 mm Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 15 If... then... installation all values are within the tol- the installation process can be started. erancerange one or more values are outside the laser analyzer must not be installed. the tolerance range OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 16: Process Purging

    Purging after To avoid damage to the parts of the gas analyzer affected by the process gas, installation process purging should be commissioned right after the gas analyzer is in- stalled. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 17: Scope Of Supply And Delivery

    OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 18: Laying Out Tools And Support Materials

    3 mm or 4 mm Slot screwdriver 6.5 mm Phillips screwdriver Phillips No. 2 Allen key 5 mm Laying out support Lay out the following support materials: materials Quantity Tools Laser alignment tool (optional) Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 19: Determining Cable Runs And Line Runs

     It must be possible to open the cable clips and line brackets to allow the cables and lines to be removed temporarily for calibration of a separate calibration set or for servicing. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 20: Installing Components

    Insertion tubes cannot be installed in conjunction with isolation flanges. Providing an overview Meaning Transmitter unit Purging flange Insertion tube Process Insertion tube Purging flange Receiver unit Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 21: Installing Insertion Tubes

    Insert the purging flange O-ring 3 between the installation flange 1 and the purging flange 4 into the groove of the purging flange. Install the purging flange 4 on the installation flange 1. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 22: Option: Installing The Isolation Flange

    Isolation flanges cannot be installed in conjunction with insertion tubes. Providing an overview Meaning Transmitter unit Purging flange Isolation flange Process Isolation flange Purging flange Receiver unit Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 23: Follow The Safety Information

    The isolation flange flat gasket is made of novaphit® SSTCTA-L, a material manufactured by Frenzelit Werke GmbH. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 24: Installing The Isolation Flange

    If necessary, loosen the lock- ing screws that have been loosely screwed in to do so 5. Insert the isolation flange flat gasket 3 between the installation flange 2 and the isolation flange 7. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 25 Tighten all fastening nuts after assembly at intervals of 24, 48 and nuts 72 hours, in order to compensate for the lingering tension release in the material of the isolation flange flat gasket. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 26: Installing The Purging Flanges

    Providing an overview Providing an overview Meaning Receiver unit Clamp Purging flange Installation flange Transmitter unit The purging flanges have a 1/4 inch Swagelok ® connection for pipes with an 8 mm outside diameter. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 27: Installing The Purging Flanges

    Step Procedure Insert the purging flange O-ring 3 into the groove of the installa- tion flange 2. Place the purging flange 4 on the installation flange 2. Screw in the flange loosely. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 28: Roughly Pre-Aligning The Purging Flanges

    Install the laser pointer on the purging flange to which the trans- mitter unit should be mounted, and the focusing screen on the purging flange to which the receiver unit should be mounted. Per- form the rough alignment. Installing the laser pointer Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 29 Place the purging flange flat gasket 2 into the groove of the purg- ing flange 4. Place the focusing screen 1 onto the purging flange 4. Fasten the focusing screen 1 using the clamp 3. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 30 Meaning Laser pointer on/off switch Laser beam Focusing screen Roughly pre-aligning the purging flanges Meaning 4 stud screws 4 fastening nuts Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 31 Switch off the laser pointer. Disassemble the laser pointer and the focusing screen. The clamps for fastening the laser alignment tool must not be used for mounting the transmitter unit or receiver unit. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 32: Connecting The Purging Lines

    Connect the purging lines as follows: Step Procedure For the purge gas supply to both purging flanges, connect a pipe with an 8 mm outside diameter to each purge gas connection (1/4 inch Swagelok ® connection). Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 33: Option: Installing The Validation Cell

    The validation cell has two 1/4 inch Swagelok ® connection for pipes with an 8 mm outside diameter. The validation cell is to be installed on the side of the transmitter unit. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 34: Installing The Validation Cell

    Fasten the validation cell 4 using the clamp 3. For the test gas supply and discharge, connect a pipe with an 8 mm outside diameter to each 1/4 inch Swagelok ® connection. Install the transmitter unit. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 35: Installing The Transmitter Unit And Receiver Unit

    Topic Page Providing an overview ..................35 Installing the transmitter unit and receiver unit ..........36 Providing an overview Providing an overview Meaning Receiver unit Clamp Purging flange Installation flange Transmitter unit OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 36: Installing The Transmitter Unit And Receiver Unit

    Attach the transmitter unit or the receiver unit using the clamp 2. Purging after To avoid damage to the parts of the gas analyzer affected by the process gas, installation process purging should be commissioned right after the gas analyzer is in- stalled. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 37: Connecting The Electrical Leads

    Connecting the power supply ................50 Providing an overview Providing an overview Meaning Receiver unit Transmitter unit Junction box Power supply T/P probes PC (can be connected temporarily) Analog and digital outputs OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 38: Observing Cable Specifications

    Max. 4 mm ; typically 2.5 mm equalization cables (AWG 8 in accordance with CSA regulations) for transmitter unit, Cable length As short as possible (minimum length 1 m) receiver unit and junction box Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 39: Protecting The Line Voltage Supply

     The way in which the supplied device is arranged must not compromise the operation of the separator.  The separator must be identified so that the assignment to the supplied device is clearly visible. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 40: Installing The Junction Box

    When mounting the junction box to a wall, use screws and wall plugs that can safely bear four times the weight of the junction box (4 x approx. 4.7 kg = approx. 19 kg). Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 41: Fitting The Cable Clips And Line Brackets

    Fit the cable clips and line brackets along the defined cable and line runs. Fitting the cable clips and line brackets OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 42: Selecting A Suitable Cable Gland

     The cable diameter must fit the diameter of the cable gland. Cable Cable gland Cable diameter Power supply 8…15 mm P/T probes 7…13 mm Analog and digital outputs 7…13 mm Receiver unit 9…17 mm Transmitter unit 9…17 mm Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 43: Leading Cables Through Cable Glands

    Screw the union nut onto the connection socket. Tightening torque: M20 power supply: 10 Nm M20 P/T probes and analog/digital outputs: 12 Nm M25 Transmitter/receiver unit connection cable: 12 Nm Close the junction box. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 44: Establishing A Protective Grounding

    The protective grounding is established via the PE conductor of protective grounding the power cable. Proceed as follows: Step Procedure Connect the PE conductor of the power cable to the PE terminal of the terminal strip. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 45: Connecting The Transmitter Unit To The Junction Box

    PE terminal. Otherwise, the wire must be insulated with shrinkdown plastic tubing and fixed to the cable harness with cable ties. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 46: Connecting The Receiver Unit To The Junction Box

    +24 V Brown Green TURU_A Yellow TURU_B Gray AUXIF_A Pink AUXIF_B AUX_IO Blue AO3 (4...20 mA) Black DO1_A Violet DO1_B Gray/pink DO2_A Blue/red DO2_B White/green T-Probe_in Brown/green T-Probe_out White/yellow P-Probe_in Brown/yellow P-Probe_out Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 47: Option: Connecting The T/P Probes To The Junction Box

    T probe in (+) 4…20-mA input T probe out (-) Analog input for dynamic temperature correction P probe in (+) 4…20-mA input P probe out (-) Analog input for dynamic pressure correction OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 48: Connecting The Analog And Digital Outputs To The Junction Box

    DO1_B DO2_A Digital output 2 DO2_B Layout of the analog The assignment of the analog and digital outputs is determined during the and digital outputs installation and configuration of the gas analyzer. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 49: Connecting Equipotential Bonding

     Receiver unit  Junction box Meaning Receiver unit Transmitter unit Junction box The potential equalization terminal prevents the cable from twist- ing during installation and prevents it from coming loose during operation. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 50: Connecting The Power Supply

    As shown in the figure, within the junction box, the primary lines and the Laying lines safely secondary lines must be laid at a distance and secured in such a way that contact between them is excluded. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 51: Gas Analyzer Start-Up

    The transmitter unit and receiver unit are in the following receiver unit condition:  The housings are not damaged  The housing covers are closed and bolted  All of the fixing screws are present. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 52 Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 53: Connecting The Supply Voltage

    Connecting the supply voltage Connecting the Connect the supply voltage. supply voltage OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 54: Recognizing The Operating Status

    The Service operating condition can be started manually by the operator. "Service" In the "Service" operating status, you can perform actions such as the follow- ing:  View the spectrum  Performing calibration  Enable the measurements log Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 55: Connecting The Pc To The Junction Box

    Connect the PC to the RJ45 adapter board using a patch cable. Connections to the terminal strip are made as follows: Terminal Signal Function ETH TX+ Ethernet TX+ ETH TX- Ethernet TX- ETH RX+ Ethernet RX+ ETH RX- Ethernet RX- OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 56: Connecting To The Instrument Software

    Start up the PC. Start up the web browser on your computer. In your web browser, type the IP address of the analyzer. The user interface of the instrument software is displayed: Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 57: Menu Structure Of The Instrument Software

    Save settings I/O verification Alignment Diagnostics Service Network settings Spectrum Calibration Instrument logging All menu illustrations in these operating instructions are examples. The actual menu display in the web browser may vary. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 58: Main Menu

    The numbers which are displayed on the left next to the log in entry field are the serial number of the transmitter unit and the manufacturing no. (F-no) of the gas analyzer. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 59: System Time Menu

    To do so, switch off the gas analyzer and power it back up again. Afterwards, check in the System time menu that the changed sys- tem time settings have been accepted. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 60: Fine Alignment Of The Purging Flanges

    The current values "Relative transmission" and "Absolute transmis- sion" appear. Click on "Align". The analyzer is now ready for fine alignment. Fine alignment of the purging flanges Meaning 4 stud screws 4 fastening nuts Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 61 If the specifications correspond to the conditions that actually apply to the process (temperature, pressure, optical path length, etc.), the analyzer can be put directly into operation after fine alignment. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 62: Alignment Menu

    Log out Operating condition of the analyzer Instrument mode Measured relative transmission Relative transmission Measured absolute transmission Absolute transmission Start fine alignment of the purging flange (see page Align Exit menu Done Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 63: Installation Procedure Menu

    Enter physical properties for the measuring environment (see page 69). Configure measuring channels (see page 70). Configure analog and digital outputs (see page 72). Check settings and save or reject (see page 74). End configuration routine. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 64: Installation - Cable Length Menu

    The length of the connection cables is provided in the analyzer data sheet. Prev Back to previous menu Cancel Cancel the process and exit the menu Next Continue to the next menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 65: Installation - Process Parameters Menu

    Select the source of the value for the pressure Pressure input FIXED: Use a manually entered process value EXTERNAL: Use the value measured using a connected P-probe AMBIENT: Use the ambient value measured by the analyzer OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 66 (for in-process measurement) NO: NH measurement without cross-sensitivity cor- rection (for calibration and validation) Back to previous menu Prev Cancel the process and exit the menu Cancel Continue to the next menu Next Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 67: Installation - Installation Flanges Menu

    Fixed temperature level the flanges Enter an offset value for the process gas pressure in Offset pressure the flanges Enter an offset value for the purging gas temperature Offset temperature in the flanges OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 68 O concentration in the flanges 0 % with dry N or 0.1 % with dry compressed air Back to previous menu Prev Cancel the process and exit the menu Cancel Continue to the next menu Next Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 69: Installation - Ambient Conditions Menu

    Height above sea level Enter the height of the installation location above sea level Prev Back to previous menu Cancel Cancel the process and exit the menu Next Continue to the next menu OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 70: Installation - Channels Menu

    Preset: 80% of the measuring range Measuring channel 3 Channel 3 Transmission - description of measuring channel 3 Output label Unit for the value at output 3 Output unit not used Output tag Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 71 Continue to the next menu The layout of the measuring channels presented in the figure and described in the table is an example. The actual layout is specific to each device and can differ from this. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 72: Installation - Analog And Digital Outputs Menu

    Value at 20 mA Enter the measured value that corresponds to a cur- rent of 20 mA Analog output 3 Analog output 3 Assign the measuring channel to the analog output Source Factory setting: Transmission Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 73 Continue to the next menu Next The layout of the outputs presented in the figure and described in the table is an example. The actual layout is specific to each device and can differ from this. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 74: Installation - Save Settings Menu

    Display of the entries and settings made during the configuration routine Save Save entries and settings End configuration routine Return to main menu Cancel Discard entries and settings End configuration routine Return to main menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 75: Maintaining And Servicing The Gas Analyzer

     Cleaning optical components (see page 77)  Cleaning housing (see page 77)  Validating analyzer (see page 91) Annual  Testing analyzer (see page 79)  Calibrating analyzer (see page 97) OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 76: Monitoring Optical Transmission

    5 mA and this means that: 20 mA  the optical transmission is sufficient.  the optical components do not require cleaning. An optical transmission > 5 % is sufficient. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 77: Checking And Cleaning The Components

     Oil-free, soft compressed air Approved and recommended cleaning agents:  For normal soiling: Conventional washing-up liquid, diluted with distilled water.  For heavy soiling: A 1:1 mixture of clean isopropyl alcohol and distilled water. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 78 Wipe the optical component dry. Clean the housing Wipe the housing surfaces with a dry or damp cloth (dampened with water). Do not use cleaning agents to clean the housing, as these may damage the seals. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 79: Testing The Analyzer

     The cables in the junction box are connected using the correct pin assignment  The cables in the cable glands are sealed so that no dust can penetrate  All of the unused cable wire ends are insulated or grounded OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 80 The O-rings and gaskets on the flanges are subject to normal wear; the level gaskets of wear can vary heavily depending on the operating conditions (especially measuring gas composition and measuring gas temperature). Therefore the condition of the O-rings and gaskets should be checked annually. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 81: Connecting To The Instrument Software

    Start up the PC. Start up the web browser on your computer. In your web browser, type the IP address of the analyzer. The user interface of the instrument software is displayed: OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 82: System Information Menu

    Number of standard settings, sample component Default settings customiza- Number of customized standard settings tion Transmitter firmware Firmware version of the transmitter unit Receiver firmware Firmware version of the receiver unit Done Exit menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 83: Instrument Status Menu

    Channels Measured value NH Measured value H Measured value transmission Transmission Exit menu Done Init Refresh the display Toggle span mode Pivot the NH validation cell in and out (see page 94) OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 84: Verification Of I/O Modules Menu

    Value for the pressure as defined in the Installation - Process parameters menu External temperature input Value for the temperature as defined in the Installation - Process parameters menu Done Exit menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 85: Diagnostics Menu

    Enter the name of a saved configuration file Browse… Search for a saved configuration file > Upload saved configuration file to the analyzer. Download Download and save user settings from the analyzer user settings file Exit menu Done OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 86: Service Menu

    Call up the Network settings menu (see page 87) Spectrum Call up the Spectrum menu (see page 88) Calibration Call up the Calibration menu (see page 105) Instrument Log Call up the Logging menu (see page 89) Done Exit menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 87: Network Settings Menu

    If the IP address has been changed, a restart of the gas analyzer is required afterwards. To do so, switch off the gas analyzer and power it back up again. analyzer Afterwards, the gas analyzer can be reached using the changed IP address. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 88: Spectrum Menu

    Log out and return to the main menu Log out View the spectrum Pause/Start Pause: View as a still image Start: View in real-time Config Customize display of the spectrum Done Exit menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 89: Logging Menu

    Current date and time in the analyzer Instrument time Date and time at which the log file was last updated Log last updated Update status display Refresh Control the logging Logging control OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 90 Name Meaning/function Start Start logging Stop Stop logging Restart Restart logging Download internal log file Download and save the log file from the analyzer Done Exit menu Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 91: Validating And Calibrating The Gas Analyzer

    Validation cell in the The diagram shows the positioning of the validation cell in the process: process Meaning Transmitter unit Validation cell Test gas supply and discharge Purging flange Process Purging flange Receiver unit OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 92 0.5 m 4 1.0 m 1.5 m 2.0 m Measuring range Optical path length resulting total concentration in the process in ppm concentration of the test gas to be used for validation in ppm Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 93: Validation

    Step Procedure During the measuring operation, depending on the application, allow nitrogen or compressed air to flow through the validation cell (flow rate approximately 2-3 l/hr). OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 94: Validate Using The Internal Validation Cell

    The Instrument status menu is displayed. The measured value now displayed for NH3 (1.61 ppm in the exam- ple) will be automatically subtracted from the measured value displayed later (see step 2). Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 95 The internal validation cell is swiveled out of the beam path. This process lasts a few seconds. Click on “Done” to exit the menu. The analyzer returns back to normal measuring mode. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 96: Checking Absorption Lines In The Spectrum

    If the position is not at 130±1, the analyzer should be inspected by an authorized service technician. Click on “Done” to exit the menu. The analyzer returns back to normal measuring mode. Make sure that purging flanges are purged. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 97: Calibrating The Gas Analyzer

    Preparing the Class I infrared laser beam invisible to the human eye. calibration set Danger of eye injuries when looking into the laser beam. Disconnect the transmitter unit from the supply voltage. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 98 Place the purging flange flat gasket in the groove of the calibration cell. Place the transmitter unit or the receiver unit onto the front side of the calibration cell. Attach the transmitter unit and the receiver unit using the clip. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 99 Open the junction box. Connect a PC to the junction box (see section Connecting a PC to the junction box (see page 55)). The analyzer is now ready for the calibration process. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 100: Temporarily Changing The Configuration

    Click on "Installation". The Installation procedure menu (see page 63) is displayed. Click on "Next" to start the configuration routine. The Installation - Cable length menu (see page 64) is displayed. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 101  100 ppm  200 ppm 40 ppm 0.175 m 160 ppm Step Procedure Leave the other settings unchanged. Click on "Next". The Installation - Installation flanges menu (see page 67) is dis- played. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 102 The Installation procedure menu is displayed. Repeat the configuration rou- tine. all settings are correct Click on "Save". The settings are saved. The Main menu is displayed. The analyzer is ready for calibra- tion. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 103: Calibration

    Unit for displaying the calibration gas con- centration (ppm for NH3) Averaging time Measuring duration for determining the average value (typically 30 s, max. 60 s.) User name (this entry is optional) Calibration signature OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 104 Observe the progress bar that displays the progress of the calibra- tion process. When the progress bar shows 100 %, the calibration process is complete. Click on "Done". The Main menu is displayed. Disassemble the calibration set. Resume measuring mode. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 105: Calibration Options Menu

    Log out and return to the main menu Log out Reset calibration to factory settings Reset user calibration Call up the Calibration settings menu (see page 106). Start user calibration Exit menu Cancel OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 106: Calibration Settings Menu

    Select the unit for displaying the calibration gas con- centration Averaging time Enter the measurement duration for determining the average value Calibration signature Enter a user name (optional) Cancel Exit menu Next Call up the Calibration menu (see page 107) Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 107: Calibration Menu

    Calibration gas concentration (NH Calibration gas concentration (H Transmission Transmission value Calibration progress Calibration process Calibration progress Progress of the calibration process Start Start calibration process Cancel Cancel the calibration process Done Exit menu OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 108: Continuing Measuring Mode

    Close the web browser. Shut down the PC. Disconnect the PC from the junction box by disconnecting the adapter board from the terminal strip. Close the junction box. The analyzer is ready to use again. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 109: Recognizing And Resolving Errors

    All others Unknown or cannot be resolved on  Inform authorized service person- your own nel of the error code OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 110: Error Messages In "Malfunction" Status

     Remove any foreign matter that may be present and that may interrupt the transmission of the laser beam  Check and correct the alignment of the transmitter unit in relation to the re- ceiver unit Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 111  If the error message persists or a sig- nificantly incorrect measurement result is displayed, contact the authorized service personnel. All others Unknown or cannot be resolved on your  Inform authorized service personnel of the error code OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 112: Gas Analyzer Shutdown

    Disconnect the transmitter unit from the junction box. Disconnect the transmitter unit from the potential equalization terminal. Disconnect the receiver unit from the junction box. Disconnect the receiver unit from the potential equalization ter- minal. Operating instruction LS4000 for NH and H OI/LS4000/NH3/H2O-EN...
  • Page 113: Disassembling The Gas Analyzer

    Remove the transmitter unit or receiver unit from the purging flange. Remove the purging flange flat gasket 1 from the groove of the purging flange. Disassemble the purging flanges and close the installation flange openings. OI/LS4000/NH3/H2O-EN Operating instruction LS4000 for NH and H...
  • Page 114 We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.