Whenever a liquid sample valve is used, an adjustable restrictor is employed on the sample outlet line
to maintain internal sample pressure and thereby keep a compressed gas liquefied.
THE LIFE OF AN LSV IS SHORTENED IF NOT USED WITHIN ITS SPECIFIED
PRESSURE AND TEMPERATURE RANGES.
OCCUR IF THE VALVE BOX TEMPERATURE EVER EXCEEDS THE LIMITS LISTED IN
TABLES 2- 4 AND 2- 5.
_________________________________________________________________________
_________________________________________________________________________
_________________________________________________________________________
_________________________________________________________________________
_________________________________________________________________________
_________________________________________________________________________
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
LSV Temperature Range
High Pressure Standard Pressure
Temperature Range
From To From To
Ambient 80_C Ambient 100_C
Ambient 150_C
Ambient 175_C
LSV Pressure Range
High Pressure Standard Pressure
Pressure Limit
PSIG PSIG
5000 1000
400
300
SVC 7- 12
HIGHLY DANGEROUS LEAKS CAN