LG 27HJ710S Owner's Manual page 5

Medical monitor
Hide thumbs Also See for 27HJ710S:
Table of Contents

Advertisement

Electro-Magnetic Immunity
This EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines)
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
NOTE
UT is THE A.C mains voltage prior to application of the test level.
IEC 60601-1-2
Compliance Level
Test Level
±6kV Contact
±6kV Contact
±8kV Air
±8kV AIR
±2kV for power
±2kV for power
supply lines
supply lines
± 1kV for input/
± 1kV for input/
output lines
output lines
±1kV differential
±1kV differential
mode
mode
±2kV common
±2kV common mode
mode
<5% UT
<5% UT
(>95% dip in UT for
(>95% dip in UT for
0.5 cycle).
0.5 cycle).
40% UT
40% UT
(60% dip in UT for
(60% dip in UT for
5 cycle)
5 cycle)
70% UT
70% UT
(30% dip in UT for
(30% dip in UT for
25cycle)
25cycle)
<5% UT
<5% UT
(<95% dip in UT)
(<95% dip in UT)
for 5s
for 5s
3 A/m
3 A/m
Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the EUT image intensifier
requires continued operation during power
mains interruptions, it is recommended that
the EUT image intensifier be powered from
an uninterruptible power supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
5