C h a p t E R 7
Setting calibration parameters
Should configuration with the Easy Setup menu not be enough to meet your needs, you will have to
calibrate your instrument and enter specific values to perform this calibration. The following pages
explain how to do this.
Setting level calibration parameters
Level calibration parameters are set under the Calibrate menu. Figure 39 illustrates these parameters
—
Figure 38 The Calibrate menu
Changing the empty/zero value
This value (probably set initially from the Easy Setup menu) represents the lowest level measurement
value in your vessel.
To set that value:
1 Select Calibrate > Level Calibration > Empty/Zero.
2 Press Edit.
3 Set the value as necessary and press OK.
By default, the empty/zero value is linked to the 4 mA value.
NOTICE
Make sure that the empty value is less than or equal to the full value.