HP 85062A Operating And Service Manual

3.5 mm electronic calibration kit

Advertisement

Quick Links

85062-90001
HP Part Number:
1995
Printed in USA
First Edition

Advertisement

Table of Contents
loading

Summary of Contents for HP HP 85062A

  • Page 1 85062-90001 HP Part Number: 1995 Printed in USA First Edition...
  • Page 2 Hewlett-Packard Company certifies that this product met its published specifications at the time shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed by the institutes calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 3: Table Of Contents

    Contents-l HP 8506ZA...
  • Page 4 HP 85062A Contents-2...
  • Page 5 Contents-3 HP 85062A...
  • Page 7: Chapter 1 General Information

    “Principles of Microwave Connector Care-Quick Reference Card”, located in the back of this manual. Refer to Chapter 7, “Replaceable Parts”, for the HP part number if another copy is needed. Or, contact your nearest HP sales office for the customer training course: “Understanding Connectors Used With Network Analyzers”.
  • Page 8: Serialnumbers

    SERIAL NUMBER PREFIX SUFFIX Figure I-1. Typical Calibration Kit Serial Number Label General Information HP 85062A...
  • Page 9: Incoming Inspection

    Table I-I. Serial Numbers of Individual Calibration Devices General Information l - 3 HP 85062A...
  • Page 11: Chapter 2 Specifications And Characteristics

    Specifications and Characteristics HP 85062A...
  • Page 12: Operating Temperature And Accuracy Enhancement

    Table 2-1. Environmental Speci$ications HP 85062A Specifications and Characteristics...
  • Page 13: Barometric Pressure And Relative Humidity

    CONDUCTORS CENTER CONDUCTORS OUTER CONDUCTOR OUTER CONDUCTOR --j MATING PLANE MATING PLANE Figure 2-l. Connector Protrusion and Recession Specifications and Characteristics HP 85062A...
  • Page 14: Connectortorque

    Table 2-2. 3.5 mm Connector Pin Depth Limits a. Approximately +2 sigma to -2 sigma of gage uncertainty based on studies performed at the factory using gages according to recommended procedures. Table 2-3. Electrical Characteristics Table 2-4. Mechanical Characteristics HP 85062A Specifications and Characteristics...
  • Page 15: Chapter 3 User Information

    User Information HP 85062A...
  • Page 17: Chapter 4 Preparation For Use - Gaging And Making Connections

    The human body almost always has some static charge. You are usually not aware of this charge because as 60 volts can destroy sensitive microcircuits. Always wear a grounding wrist strap connected to a conductive bench mat when working near this and other sensitive equipment. Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 18: Visualinspection

    Flat contact HP 85062A Preparation For Use - Gaging and Making Connections...
  • Page 19: Cleaningconnectors

    If you must use a solvent, use only isopropyl alcohol. Use the least amount of alcohol possible, and avoid wetting any plastic parts in the connectors. 5. w. Inspect the connector again under a magnifyin, o residues remain. Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 20: Gaging Devices In This Kit

    Averaging a minimum of three readings, each taken after a quarter-turn rotation of the gage, reduces Note measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis. HP 85062A Preparation For Use - Gaging and Making Connections...
  • Page 21: Connector Gages

    Figure 4-l. Typical 3.5 mm Connector Gage HP 85062A Preparation For Use - Gaging and Making Connections...
  • Page 22 Clean both of these carefully and check the zero setting again. HP 85062A Preparation For Use - Gaging and Making Connections...
  • Page 23: Gaging Techniques

    Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 24 Figure 4-2. Gaging a 3.5 mm Male Connector Using a Push-on Type Gage Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 25: Male 3.5 Mm (Screw-On Type Gage)

    Figure 4-3. Gaging a 3.5 mm Male Connector Using a Screw-on Type Gage Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 26: Female 3.5 Mm (Push-On Type Gage)

    Figure 4-4. Gaging a 3.5 mm Female Connector Using a Push-on Type Gage HP 85062A 4-10 Preparation For Use - Gaging and Making Connections...
  • Page 27: Female 3.5 Mm (Screw-On Type Gage)

    Figure 4-5. Gaging a 3.5 mm Female Connector Using a Screw-on Type Gage 4-l 1 Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 28: Making Connections

    Also, remember that the setting of the torque wrench can change with use. Re-calibrate at a schedule appropriate to the amount of use. HP 85062A 4-12 Preparation For Use - Gaging and Making Connections...
  • Page 29 HANDLE BEGINS TO BREAK (CORRECT POINT TO STOP) HOLD HERE HANDLE FULLY BROKEN ..-_.. .’ (NOT RECOMMENDED) DIRECTION Figure 4-6. Using the Torque Wrench 4-l 3 Preparation For Use - Gaging and Making Connections HP 85062A...
  • Page 31: What Recertification Provides

    In some cases, a kit may be not used for some time after the actual recertification date. The recertification interval should begin on the date the kit is first used. Performance Verification HP 85062A...
  • Page 32: Where To Send A Kit For Recertification

    HP 85062A Performance Verification...
  • Page 33: Chapter 6 Troubleshooting

    Troubleshooting HP 85062A...
  • Page 34 Figure 6-l. Troubleshooting Flow Diagram HP 85062A Troubleshooting...
  • Page 35 HP 85062A calibration kit and for or.h- Table 7-1. Replaceable Pans DESCRIPTION HP PART NUMBER Insertable (Standard) 85062-60002 85062-60004 Non-Insertable Male (Option OOM) 85062-60006 Non-Insertable Female (Option OOF) Low Band Calibration Modules Insertable (Option 001) 85062-60001 85062-60003 Non-Insertable Male (Option 001 and Option OOM)
  • Page 36 Operating and Service Manual, Storage Case, Specification / Verification Program Disk. Notes: Low band module shown is included only with Option 001. Serial numbers are located on opposite side of modules. Figure 7-1. HP 85062A, Option 001, Replaceable Parts Replaceable Parts HP 85062A...
  • Page 37 Operating and Service Manual, Storage Case, Specification / Verification Program Disk. Notes: Low band module shown is included only with Option 001. Serial numbers are located on opposite side of modules. Figure 7-2. HP 85062A, Option 001 and Option OOM, Replaceable Parts HP 85062A Replaceable Parts...
  • Page 38 Operating and Service Manual, Storage Case, Specification /Verification Program Disk. Notes: Low band module shown is included only with Option 001. Serial numbers are located on opposite side of modules. Figure 7-3. HP 85062A, Option 001 and Option OOF, Replaceable Parts HP 85062A Replaceable Parts...
  • Page 39 For additional connector care instruction, contact your local Hewlett-Packard Sales and Service Office about course numbers HP 85050A+24A and HP 85050A+24D. See the reverse side of this notice for quick reference tips about connector care.
  • Page 40 Principles of Microwave Connector Care-- Quick Reference Card Handling and Storage Do Not Keep connectors clean Touch mating-plane surfaces . Extend sleeve or connector nut Set connectors contact-end down Use plastic end caps during storage Visual Inspection Do Not Inspect all connectors carefully before every connection Use a damaged connector--ever Look for metal particles, scratches,...