Sony MCC-1000MD Instructions For Use Manual page 11

Hide thumbs Also See for MCC-1000MD:
Table of Contents

Advertisement

Caution
When you dispose of the unit or accessories,
you must obey the laws in the relative area
or country and the regulations in the relative
hospital regarding environmental pollution.
Warning on power
connections
Use a proper power cord for your local
power supply.
1. Use the approved Power Cord (3-core
mains lead) / Appliance Connector /
Plug with earthing-contacts that
conforms to the safety regulations of
each country if applicable.
2. Use the Power Cord (3-core mains lead)
/ Appliance Connector / Plug
conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the
above Power Cord / Appliance Connector /
Plug, please consult a qualified service
personnel.
Warning on power
connections for medical use
Customers in the U.S.A. and Canada should
use the following type of power cord.
Customers in other countries or regions
should use the power cord prescribed by
their country or region.
Plug type
Cord type
Minimum rating for
plug and appliance
couplers
Safety approval
* Note: Grounding reliability can only be
achieved when the equipment is
connected to an equivalent receptacle
marked "Hospital Only" or "Hospital
Grade".
Warning
The apparatus shall not be exposed to
dripping or splashing. No objects filled with
liquids, such as vases, shall be placed on the
apparatus.
Caution
When installing, ensure the following space
around the periphery of the unit, taking
ventilation and servicing into consideration.
– Rear side: 10 cm (4 in.) or more
– Left/Right sides: 10 cm (4 in.) or more
– Top side: 10 cm (4 in.) or more
Caution
Do not use the device in a MR (Magnetic
Resonance) environment.
It may cause a malfunction, fire, and
unwanted movement.
11
U.S.A. and
Canada
HOSPITAL
*
GRADE
Min. Type SJT
Min. 18 AWG
10 A / 125 V
UL Listed and CSA