HP 11757B Installation And Calibration Manual page 39

Multipath fading simulator
Hide thumbs Also See for 11757B:
Table of Contents

Advertisement

HP 11757B
Procedure
Note
Note
Note
allow the user to run the test in either the 70 or 140
MHz frequency band.
The Multipath Fading Simulator requires one hour
of warm-up prior to performing these tests and
adjustments.
1. Using the
or
485
495
over the desired adjustment.
2. Highlight TEST 04: COMPLETE
PERFORMANCE TESTS (01, 02, 03).
3. Press the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Item
to run.
This procedure is to be used whenever any kind of repair
or alteration is made to the HP 11757B, even if it only
involves loosening and re-tightening a cable. Figure 1-11,
\Adjustment Procedure Flowchart", illustrates how the
adjustment procedure should be performed.
4. Check the PASS/FAIL STATUS printed at the
bottom of each printout page.
The Standard model and Option 140 provides 4
printouts plus a summary sheet.
The Option 147 provides 8 printouts plus a summary
sheet.
5. Check the PASS/FAIL STATUS printed at the
bottom of each page and the summary sheet.
6. If all the statuses are \Passed", the tests are
complete. Press the
Installation and Calibration
keys, position the highlight bar
softkey. The tests will start
softkey.
NNNNNNNNNNNNNN
Exit
31