Download Print this page

Philips IntelliVue MP40 Hardware Upgrade Installation Manual page 19

Patient monitor
Hide thumbs Also See for IntelliVue MP40:

Advertisement

Installation Instructions
2
Unplug the Navigation Point connector, the MSL cable, the power suppy cable, the speaker cable
and the ECG Out/Alarm LED cable from the main board.
ECG OUT/Alarm LED
Cable
(not connected in this picture)
MSL flat ribbon
cable
To remove the flat ribbon cables with red connectors (e.g. Navigation Point cable) simply pull directly
N O TE
on the cable. The MSL, power supply and speaker cables have different security locks and must be
unlocked in order to be removed.
Remove the nine screws, pull the main board out of the rack connector and remove the board by
3
moving it towards the bottom of the monitor.
1
1a
2
1a
2
2
1
2
3
5 Installing E12
Power Supply Cable
Speaker Cable
Navigation Point
Connector
(not connected in this picture)