Download Print this page

Verification Of Upgrade - Philips IntelliVue MP40 Hardware Upgrade Installation Manual

Patient monitor
Hide thumbs Also See for IntelliVue MP40:

Advertisement

Verification of Upgrade

7
After an upgrade you must test the monitor to ensure that the unit has full functionality.
If an upgrade has been performed where the unit has not been opened (i.e. front and back half of the
monitor have not been separated):
Perform Power On Test block.
1
If an upgrade has been performed where the unit has been opened (i.e. front and back half of the
monitor have been separated):
Perform Power On Test block.
1
2
Perform Safety Test block.
For specific performance tests or documentation requirements see the IntelliVue MP40/MP50 service
guide.
7