Troubleshooting - LG TEC Multi Instruction Manual

Table of Contents

Advertisement

If your device does not seem to be operating correctly, refer
to the chart below to determine what may be wrong. Should
none of these measures correct the problem, the device
should be serviced.
Problem
Possible Cause
Displays fail
Battery contact
to light up
failure
Electrodes
Stimulation
1. Dried out or
weak
contaminated
2. Placement
Lead wires
Old/worn/damaged
Intensity is too high
Stimulation is
uncomfortable
Electrodes are
too close together
Damaged or worn
electrodes or lead
wires
Electrode active
area size is too
small.
Mayn't operate the
device according
to the manual.
Intermittent
Lead wires
output

8. TROUBLESHOOTING

30
Solution
1. Try fresh batteries.
2. Ensure batteries are
inserted correctly.
Check the following
contacts:
• All contacts are in place.
• All contacts are not broken.
Replace and
re-connect
Replace
Decrease intensity.
Reposition the electrodes.
Replace.
Replace electrodes with ones
that have an active area no
2
less than 16.0cm (4cm*4cm).
Please check the manual
before use.
1. Verify connection is
secure. Insure firmly.
2. Turn down the intensity.
Rotate lead wires in socket
90°. If still intermittent,
replace lead wire.