HP 37718A Installation And Verification Manual page 94

Communications performance analyzer
Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

N OTE
Performance Tests
PDH/DSn Transmitter Output
4 Set the
TRANSMIT
shown opposite.
5 Select CHAN 1 on the oscilloscope
and set to 1MΩ input impedance.
Load the Telecom Mask
6 On the oscilloscope, use the
following key sequence to select and load the required Telecom Mask:
a/ SHIFT (Blue Key) APPLICATION.
b/
TELECOM MASK/MASK
c/ Use ENTRY/MEASURE control to select DS1_NEW in the
highlighted MSK FILES window.
d/
LOAD SETUP 8 M1M2
Compare Pulse with Mask
7 On the oscilloscope, use the following key sequence to compare the
pulse with the mask:
a/ RECALL 8, (Ensure CH1 Impedance is reset to 1M
TRIGGER,
to
).
Trig'd
b/ DEFINE MEAS
c/
softkey
COMPARE
d/
softkey
TEST ON
e/
softkey
M1M2
f/
AFTER FAIL CONTINUE
g/ RUN
Using the balanced to unbalanced converter reduces the pulse amplitude
by a factor of 1.266. This will cause the comparison with the mask to fail.
To overcome this select CHAN 1 ON and reduce the Volts/Division setting
(highlighted on the softkey display). Reduced Volts/Division = highlighted
value/1.266. The new value can be entered using the keypad
8 The oscilloscope will compare the positive pulse with the mask limits.
A PASS message should appear on the Oscilloscope.
3-26
display as
softkey.
softkey.
SETUP and set
softkey
. Select
softkey
PATTERN TRIGGER