Interconnections; Power Requirements; Line Voltage And - HP 4085M Operation Manual

Hide thumbs Also See for 4085M:
Table of Contents

Advertisement

Section
11
2-S.
I N TERCO N N ECTIO N S
To construct the system shown in Figure 1 - 2 , interconnect the instrum ents as .
2-9.
shown in Figure 2-5. The required instruments and cables are listed iri the figure. The
connection procedure is as follows:
Interconnect the 4084A
1.
2. Attach the 1 6 07SA
then tighten the two, thumb screws.
3. Connect each S M U terminal on the rear of the 4145A
terminal on the 1 6 078A. Use the f�rnished triaxial cablesC!).
4. Connect the VM and VS terminals of the 4 1 45A to the VM PORTs and VS PORTs of
the 40S5A with the furnished B N C cablesC!) .
5. Connect the 24-pin Shorting Connector 0) to the System Cable connector of the
41 45A.
6. Interconnect the 40S4A, 4145A, and computer (internal HP-IB interface) with the
furnished H P-IB cablesG).
7. Set the HP-IB Addresses of the 40S4A and 4145A as listed below :
4084A: 2 2
4145A: 17
If other HP-IB addresses are used, or if an HP-I B interface card is used instead of the
computer's internal HP-IB interface, the furnished software m ust be m odified, as
described in paragraph 3-32.
2-1 0. POWER REQUIRE M E N TS
2-1 1. The 4085M requires a power source of 1 00, 120, 220 Volts ac ±1 0%. or 240 Volts ac
+5%-1 0%, 48 to 66Hz single phase; power consumption is 1 3 0VA maximum.
IF THE 40S4A IS TO BE E N ERGIZED VIA A N EXTERNAL AUTOTRA NS­
FORMER
CO M M O N TER MINAL IS CO N N ECTED TO THE N EUTRAL POLE OF THE
POWER SUPPLY.
2-12 . Line Voltage and Fuse Selection
BEFORE
SELECTOR SWITCH O N THE REAR PANEL IS CORRECTLY 'SET FOR
THE LI NE VOLTAGE USED.
2-1 3. Figure-2-6 shows the various settings of the line voltage selector switches and lists,
the proper fuses for different line voltages. When the 4084A
its line voltage selector switch is correctly set and the proper fuse is installed for the line
voltage used at the shipment's destination.
2- 4
0)
and 4085A
®
to the 40S5A conn
U N IT
FOR
VOLTAGE
TUR NI N G
O N
THE
(j)
with the furnished control cable
'
ectors labeled TO 1 607SA ADAPTER, and
WARNING
REDUCTIO N ,
BE
CAUTION
40S4A,
VERIFY
THAT
is
M odel 40S5M
.
0.
CD
to the corresponding
.
SURE
THAT THE
THE VOLTAGE
shipped from the factory,