Routine Maintenance; Cleaning The Chamber And Inner Attachments - Sanyo MCO-19AIC(UV) Instruction Manual

Co2 incubator
Hide thumbs Also See for MCO-19AIC(UV):
Table of Contents

Advertisement

ROUTINE MAINTENANCE

Cleaning the chamber and Inner attachments

When using the Incubator for the first time or when performing H
Chamber Attachments on the following page and remove and clean the inner attachments.
WARNING
Before performing any repairs or maintenance, turn OFF the power supply switch and unplug the
Incubator. Failure to do say may result in electric shock or injury.
Be careful not to inhale chemicals, vapors or aerosols when cleaning the Incubator. Doing so may
be harmful to health.
CAUTION
Wear rubber gloves when performing maintenance on the chamber. Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
Note:
Be careful not to damage the humidifying pan water level sensor or the UV lamp in the chamber duct (if
the MCO-19AIC(UV) or the optional MCO-19UVS is installed).
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
1. Turn OFF the power to the Incubator.
2. Open the outer and inner doors and pull out the trays. (See Fig. 1.)
3. Lift the front of the tray supports and pull them out. (See Fig. 2.)
4. Lift the humidifying pan cover off from the pins on the rear side. (See Fig. 3.)
Fig. 1
O
decontamination, refer to Removing
2
2
Fig. 2
127
Fig. 3
36

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-19aic

Table of Contents