Sanyo MCO-19AIC(UV) Instruction Manual page 48

Co2 incubator
Hide thumbs Also See for MCO-19AIC(UV):
Table of Contents

Advertisement

The buzzer can be stopped by pressing the BUZZER Key, but remote alarms will not stop. In addition, a
buzzer that sounds because of the upper limit temperature alarm cannot be stopped by pressing the
BUZZER Key.
Err01 recovery is achieved automatically by correctly connecting the gas and pressing the BUZZER Key
to stop the buzzer. In addition, when the MCO-21GC is installed and the gas supply is switched to the
reserve gas cylinder, press the BUZZER Key and replace the gas cylinder.
If an error from Err05 to Err18 occurs (except for the upper limit temperature alarm in Err13 to Err16),
then consult a Sanyo sales representative or agent.
Table 2: Alarm and Safety Functions for H
Alarm and safety
function
System check error
The H
O
2
at start of H
O
connected.
2
2
decontamination
There is no H
the H
O
2
failed (or is disconnected).
A door is not closed.
Failure during H
O
The volume of H
2
2
decontamination
is abnormal.
Power was interrupted during
H
O
decontamination.
2
2
Failure during H
O
The UV lamp failed during
2
2
gas analysis by UV
H
O
gas analysis by UV.
2
2
Power was interrupted during
H
O
gas analysis by UV.
2
2
Table 3: Alarm and Safety Functions for Automatic Standard Gas Calibration
Alarm and safety
function
System check error
The gas pressure for the
at start of automatic
standard gas calibration CO
standard gas
gas line is less than the
calibration
specified pressure.
Failure during
The gas pressure for the
automatic standard
standard gas calibration CO
gas calibration
gas line is less than the
specified pressure.
Note
For models with electric locking, an unlock key is provided in order to unlock the door during a power
interruption or in case the electric lock fails. Always store this key in a safe place. It is recommended that
you make a note of the key symbol and number in case the key is lost.
WARNING
Do not use the unlock key to unlock the door during H
analysis by UV. Doing so may cause harm to health from H
47
O
2
Condition
Generator is not
H2O2 Decon System Error is displayed
2
in the screen center.
Err31:H2O2 Unit No Connect is
displayed in the message display field.
O
solution or
H2O2 Decon System Error is displayed
2
2
level sensor has
in the screen center.
2
Err32: H2O2 Level is Low is displayed in
the message display field.
H2O2 Decon System Error displayed in
the screen center.
Err33: Door is Open is displayed in the
message display field.
O
generated
Err34:H2O2 Volume is displayed in the
2
2
message display field.
After power is restored, Err35: Power
Failed is displayed in the screen
message display field.
After completion, Decon Failed is
displayed in the screen center.
Err36: UV Lamp Failure is displayed in
the message display field.
After power is restored, Err37: Power
Failed is displayed in the message
display field.
Condition
Std Gas Calib Error is displayed in the
screen center.
2
Err41: CO2 Std Gas Empty is displayed
in the message display field.
Std Gas Calib Error is displayed in the
screen center.
2
Err42: CO2 Std Gas Empty is displayed
in the message display field.
Decontamination
2
Display
Display
O
decontamination or during H
2
2
O
gas leakage.
2
2
138
Buzzer
Safety operation
-----
Decontamination mode is
cancelled.
-----
Decontamination mode is
cancelled.
-----
Decontamination mode is
cancelled.
Intermittent
Moves to UV analysis.
tone
Remote alarm operates.
Intermittent
Electrically locked during power
tone
interruption. After power is
restored, moves to UV analysis.
Remote alarm operates.
Intermittent
UV analysis time is extended.
tone
Remote alarm operates.
-----
Electrically locked during power
interruption. After power is
restored, UV analysis is
repeated.
Buzzer
Safety operation
Intermittent
The standard gas calibration
tone
mode is cancelled.
Remote alarm operates.
Intermittent
Automatic calibration is
tone
cancelled.
Remote alarm operates.
O
2
gas
2