The monitor will display a Leads off message in an input overload condition, or
●
upon disconnection of electrode leadwires.
ECG points to note
This equipment is suitable for use in the presence of electrosurgery, as tested
●
according to IEC 60601-2-49 clause 202.6.2.101 Electrosurgery interference.
For information on materials used in accessories and their biocompatibility, refer
●
to the instructions for use in the accessory package.
Defibrillator discharge may affect the ECG measurement. Refer to the supplemental
●
information provided for details regarding the recovery time from defibrillator
discharge.
Pre-gelled ECG electrodes are recommended. Check the expiration date.
●
Make sure the electrode gel is moist.
●
Make sure the electrodes have good skin contact.
●
Replace all electrodes at least every 24 to 48 hours.
●
Select the Deactivate ECG Leads Off option to remove a Leads off message from
●
the display when a cable is off.
ECG measurement setup
ECG equipment to patient connection with CARESCAPE ONE
2094480-001
HEART RATE ALARM INTERFERENCE. Poor cable positioning
or improper electrode preparation may cause line isolation
monitor transients to resemble actual cardiac waveforms and
thus inhibit heart rate alarms. To minimize this problem, follow
proper electrode placement and cable positioning guidelines
provided with this device.
The patient's skin may become irritated after prolonged
contact with electrode gel or adhesive.