Ecg Cautions; Ecg Measurement Limitations; Ecg Points To Note; Ecg Measurement Setup - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
WARNING

ECG cautions

CAUTION

ECG measurement limitations

The monitor will display a Leads off message in an input overload condition, or
upon disconnection of electrode leadwires.

ECG points to note

This equipment is suitable for use in the presence of electrosurgery, as tested
according to IEC 60601-2-49 clause 202.6.2.101 Electrosurgery interference.
For information on materials used in accessories and their biocompatibility, refer
to the instructions for use in the accessory package.
Defibrillator discharge may affect the ECG measurement. Refer to the supplemental
information provided for details regarding the recovery time from defibrillator
discharge.
Pre-gelled ECG electrodes are recommended. Check the expiration date.
Make sure the electrode gel is moist.
Make sure the electrodes have good skin contact.
Replace all electrodes at least every 24 to 48 hours.
Select the Deactivate ECG Leads Off option to remove a Leads off message from
the display when a cable is off.

ECG measurement setup

ECG equipment to patient connection with CARESCAPE ONE

2094480-001
HEART RATE ALARM INTERFERENCE. Poor cable positioning
or improper electrode preparation may cause line isolation
monitor transients to resemble actual cardiac waveforms and
thus inhibit heart rate alarms. To minimize this problem, follow
proper electrode placement and cable positioning guidelines
provided with this device.
The patient's skin may become irritated after prolonged
contact with electrode gel or adhesive.
CARESCAPE ONE
ECG
99